This site displays a prototype of a “Web 2.0” version of the daily
Federal Register. It is not an official legal edition of the Federal
Register, and does not replace the official print version or the official
electronic version on GPO’s govinfo.gov.
The documents posted on this site are XML renditions of published Federal
Register documents. Each document posted on the site includes a link to the
corresponding official PDF file on govinfo.gov. This prototype edition of the
daily Federal Register on FederalRegister.gov will remain an unofficial
informational resource until the Administrative Committee of the Federal
Register (ACFR) issues a regulation granting it official legal status.
For complete information about, and access to, our official publications
and services, go to
About the Federal Register
on NARA's archives.gov.
The OFR/GPO partnership is committed to presenting accurate and reliable
regulatory information on FederalRegister.gov with the objective of
establishing the XML-based Federal Register as an ACFR-sanctioned
publication in the future. While every effort has been made to ensure that
the material on FederalRegister.gov is accurately displayed, consistent with
the official SGML-based PDF version on govinfo.gov, those relying on it for
legal research should verify their results against an official edition of
the Federal Register. Until the ACFR grants it official status, the XML
rendition of the daily Federal Register on FederalRegister.gov does not
provide legal notice to the public or judicial notice to the courts.
Rule
Enter a search term or FR citation e.g.
88 FR 38230 FR 78782024-13208USDA09/05/24RULE0503-AA39SORN
Choosing an item from
full text search results
will bring you to those results. Pressing enter in the search box
will also bring you to search results.
Choosing an item from
suggestions
will bring you directly to the content.
Medical Devices; Hematology and Pathology Devices; Classification of the Container System for the Processing and Storage of Red Blood Cell Components Under Reduced Oxygen Conditions
This order is effective November 9, 2023. The classification was applicable on September 15, 2023.
Table of Contents
Enhanced Content - Table of Contents
This table of contents is a navigational tool, processed from the
headings within the legal text of Federal Register documents.
This repetition of headings to form internal navigation links
has no substantive legal effect.
Comments are no longer being accepted.
See DATES for details.
Regulations.gov Data
Enhanced Content - Regulations.gov Data
FederalRegister.gov retrieves relevant information about this document
from Regulations.gov to provide users with additional context. This
information is not part of the official Federal Register document.
Medical Devices; Hematology and Pathology Devices; Classification of the Container System For The Processing And Storage Of Red Blood Cell Components Under Reduced Oxygen Conditions
Document page views are updated periodically throughout the day and are
cumulative counts for this document. Counts are subject to sampling,
reprocessing and revision (up or down) throughout the day.
Page views
2,173
as of
08/13/2026 at 4:15 am EDT
Other Formats
Enhanced Content - Other Formats
This document is also available in the following formats:
This PDF is FR Doc. 2023-24717 as it appeared on Public Inspection on
11/08/2023 at 8:45 am.
It was viewed
38
times while on Public Inspection.
If you are using public inspection listings for legal research, you
should verify the contents of the documents against a final, official
edition of the Federal Register. Only official editions of the
Federal Register provide legal notice of publication to the public and judicial notice
to the courts under 44 U.S.C. 1503 & 1507.
Learn more here.
Published Document: 2023-24717 (88 FR 77198)
This document has been published in the Federal Register. Use the PDF linked in the document sidebar for the official electronic format.
AGENCY:
Food and Drug Administration, Department of Health and Human Services (HHS).
ACTION:
Final amendment; final order.
SUMMARY:
The Food and Drug Administration (FDA, Agency, or we) is classifying the container system for the processing and storage of Red Blood Cell components under reduced oxygen conditions into class II (special controls). The special controls that apply to the device type are identified in this order and will be part of the codified language for the classification of the container system for the processing and storage of Red Blood Cell components under reduced oxygen conditions. We are taking this action because we have determined that classifying the device into class II (special controls) will provide a reasonable assurance of safety and effectiveness of the device. We believe this action will also enhance patients' access to beneficial innovative devices.
DATES:
This order is effective November 9, 2023. The classification was applicable on September 15, 2023.
FOR FURTHER INFORMATION CONTACT:
Karen Fikes, Center for Biologics Evaluation and Research, Food and Drug Administration, 10903 New Hampshire Ave., Bldg. 71, Rm. 7301, Silver Spring, MD 20993-0002, 240-402-7911.
SUPPLEMENTARY INFORMATION:
I. Background
Upon request, FDA has classified the container system for the processing and storage of Red Blood Cell components under reduced oxygen conditions as class II (special controls), which we have determined will provide a reasonable assurance of safety and effectiveness. The automatic assignment of class III occurs by operation of law and without any action by FDA, regardless of the level of risk posed by the new device. Any device that was not in commercial distribution before May 28, 1976, is automatically classified as, and remains within, class III and requires premarket approval unless and until FDA takes an action to classify or reclassify the device (see 21 U.S.C. 360c(f)(1)). We refer to these devices as “postamendments devices” because they were not in commercial distribution prior to the date of enactment of the Medical Device Amendments of 1976, which amended the Federal Food, Drug, and Cosmetic Act (FD&C Act).
FDA may take a variety of actions in appropriate circumstances to classify or reclassify a device into class I or II. We may issue an order finding a new device to be substantially equivalent under section 513(i) of the FD&C Act (see 21 U.S.C. 360c(i)) to a predicate device that does not require premarket approval. We determine whether a new device is substantially equivalent to a predicate device by means of the procedures for premarket notification under section 510(k) of the FD&C Act (21 U.S.C. 360(k)) and part 807 (21 CFR part 807).
FDA may also classify a device through “De Novo” classification, a common name for the process authorized under section 513(f)(2) of the FD&C Act; (see also 21 CFR part 860, subpart D). Section 207 of the Food and Drug Administration Modernization Act of 1997 (Pub. L. 105-115) established the first procedure for De Novo classification. Section 607 of the Food and Drug Administration Safety and Innovation Act (Pub. L. 112-144) modified the De Novo application process by adding a second procedure. A device sponsor may utilize either procedure for De Novo classification.
Under the first procedure, the person submits a 510(k) for a device that has not previously been classified. After receiving an order from FDA classifying the device into class III under section 513(f)(1) of the FD&C Act, the person then requests a classification under section 513(f)(2).
Under the second procedure, rather than first submitting a 510(k) and then a request for classification, if the person determines that there is no legally marketed device upon which to base a determination of substantial equivalence, that person requests a classification under section 513(f)(2) of the FD&C Act.
Under either procedure for De Novo classification, FDA is required to classify the device by written order within 120 days. The classification will be according to the criteria under section 513(a)(1) of the FD&C Act. Although the device was automatically placed within class III, the De Novo classification is considered to be the initial classification of the device.
When FDA classifies a device into class I or II via the De Novo process, the device can serve as a predicate for future devices of that type, including for 510(k)s (see section 513(f)(2)(B)(i) of the FD&C Act). As a result, other device sponsors do not have to submit a De Novo request or premarket approval application to market a substantially equivalent device (see section 513(i) of the FD&C Act, defining “substantial equivalence”). Instead, sponsors can use the 510(k) process, when necessary, to market their device.
II. De Novo Classification
On January 5, 2022, FDA received Hemanext, Inc.'s request for De Novo classification of the Hemanext One. FDA reviewed the request in order to classify the device under the criteria for classification set forth in section 513(a)(1) of the FD&C Act.
We classify devices into class II if general controls by themselves are insufficient to provide reasonable assurance of safety and effectiveness, but there is sufficient information to establish special controls that, in combination with the general controls,
( printed page 77199)
provide reasonable assurance of the safety and effectiveness of the device for its intended use (see 21 U.S.C. 360c(a)(1)(B)). After review of the information submitted in the request, we determined that the device can be classified into class II with the establishment of special controls. FDA has determined that these special controls, in addition to the general controls, will provide reasonable assurance of the safety and effectiveness of the device.
Therefore, on September 15, 2023, FDA issued an order to the requester classifying the device into class II. In this final order, FDA is codifying the classification of the device by adding 21 CFR 864.9115.[1]
We have named the generic type of device container system for the processing and storage of Red Blood Cell components under reduced oxygen conditions, identified as a device intended for medical purposes that is used to process and store Red Blood Cell components and reduce oxygen levels in the storage environment.
FDA has identified the following risks to health associated specifically with this type of device and the measures required to mitigate these risks in table 1.
Table 1—Container System for the Processing and Storage of Red Blood Cell Components Under Reduced Oxygen Conditions Risks to Health and Mitigation Measures
Identified risks to health
Mitigation measures
Toxicity that can result from contact of the component materials of the device with the red blood cells or patient's body
Biocompatibility evaluation.
Toxicity of leached materials, or residual chemical sterilant, when in contact with red blood cells or transfused to patient
Extractables and leachables testing.
Infection
Sterilization validation; Endotoxin testing; and Container closure evaluation.
Transfusion of poor quality red blood cells because of inadequate storage conditions or device malfunction
Nonclinical and clinical studies; Shelf-life testing; and Performance testing.
Blood exposure because of device malfunction
Performance testing.
Transfusion of poor quality red blood cells due to processing of Red Blood Cells components collected from donors with hemoglobin S
Labeling.
FDA has determined that special controls, in combination with the general controls, address these risks to health and provide reasonable assurance of safety and effectiveness. For a device to fall within this classification, and thus avoid automatic classification in class III, it would have to comply with the special controls named in this final order. The necessary special controls appear in the regulation codified by this order. This device is subject to premarket notification requirements under section 510(k) of the FD&C Act. At the time of classification, container system for the processing and storage of Red Blood Cell components under reduced oxygen conditions is for prescription use only. Prescription devices are exempt from the requirement for adequate directions for use for the layperson under section 502(f)(1) of the FD&C Act (21 U.S.C. 352(f)(1)) and 21 CFR 801.5, as long as the conditions of 21 CFR 801.109 are met.
III. Analysis of Environmental Impact
The Agency has determined under 21 CFR 25.34(b) that this action is of a type that does not individually or cumulatively have a significant effect on the human environment. Therefore, neither an environmental assessment nor an environmental impact statement is required.
IV. Paperwork Reduction Act of 1995
This final order establishes special controls that refer to previously approved collections of information found in other FDA regulations and guidance. These collections of information are subject to review by the Office of Management and Budget (OMB) under the Paperwork Reduction Act of 1995 (44 U.S.C. 3501-3521). The collections of information in 21 CFR part 860, subpart D, regarding De Novo classification have been approved under OMB control number 0910-0844; the collections of information in 21 CFR part 814, subparts A through E, regarding premarket approval, have been approved under OMB control number 0910-0231; the collections of information in part 807, subpart E, regarding premarket notification submissions, have been approved under OMB control number 0910-0120; the collections of information in 21 CFR part 820, regarding quality system regulation, have been approved under OMB control number 0910-0073; and the collections of information in 21 CFR part 801, regarding labeling, have been approved under OMB control number 0910-0485.
Therefore, under the Federal Food, Drug, and Cosmetic Act and under authority delegated to the Commissioner of Food and Drugs, 21 CFR part 864 is amended as follows:
PART 864—HEMATOLOGY AND PATHOLOGY DEVICES
1. The authority citation for part 864 continues to read as follows:
Container system for the processing and storage of Red Blood Cell components under reduced oxygen conditions.
(a)
Identification.
A container system for the processing and storage of Red Blood Cell components under reduced oxygen conditions is a device intended for medical purposes that is used to process and store Red Blood Cell components and reduce oxygen levels in the storage environment.
(b)
Classification.
Class II (special controls). The special controls for this device are:
( printed page 77200)
(1) The intended use of the device must specify:
(i) The Red Blood Cell components that can be processed and stored including acceptable anticoagulants and additive solutions;
(ii) The hold time after Red Blood Cell component collection;
(iii) The processing capacity (volume) of the device; and
(iv) The storage temperature and dating period of processed Red Blood Cell components.
(2) Studies must demonstrate that the device is biocompatible and include detailed documentation of the biocompatibility evaluation.
(3) Performance testing and nonclinical studies must include a detailed study of leached materials extracted under conditions similar to clinical usage of the device, and a toxicologic risk assessment of those extracted or leached materials.
(4) Performance testing must support sterility of the device and include sterilization validation, endotoxin testing, and container closure integrity evaluation.
(5) Nonclinical and clinical studies must include evaluation of red blood cell quality throughout the duration of storage based on in vitro and in vivo studies, including hemolysis and red blood cell survival and recovery.
(6) Performance studies must include:
(i) Detailed documentation of functional and mechanical testing, including evaluation of oxygen and, if applicable, carbon dioxide levels during Red Blood Cell components storage; and
(ii) Detailed documentation of device shelf-life testing demonstrating continued sterility, package integrity, and functionality over the identified shelf life.
(7) The labeling must include a contraindication against processing Red Blood Cell components collected from donors with hemoglobin S.
Dated: November 3, 2023.
Lauren K. Roth,
Associate Commissioner for Policy.
Footnotes
1.
FDA notes that the “ACTION” caption for this final order is styled as “Final amendment; final order,” rather than “Final order.” Beginning in December 2019, this editorial change was made to indicate that the document “amends” the Code of Federal Regulations. The change was made in accordance with the Office of Federal Register's (OFR) interpretations of the Federal Register Act (44 U.S.C. chapter 15), its implementing regulations (1 CFR 5.9 and parts 21 and 22), and the Document Drafting Handbook.