This site displays a prototype of a “Web 2.0” version of the daily
Federal Register. It is not an official legal edition of the Federal
Register, and does not replace the official print version or the official
electronic version on GPO’s govinfo.gov.
The documents posted on this site are XML renditions of published Federal
Register documents. Each document posted on the site includes a link to the
corresponding official PDF file on govinfo.gov. This prototype edition of the
daily Federal Register on FederalRegister.gov will remain an unofficial
informational resource until the Administrative Committee of the Federal
Register (ACFR) issues a regulation granting it official legal status.
For complete information about, and access to, our official publications
and services, go to
About the Federal Register
on NARA's archives.gov.
The OFR/GPO partnership is committed to presenting accurate and reliable
regulatory information on FederalRegister.gov with the objective of
establishing the XML-based Federal Register as an ACFR-sanctioned
publication in the future. While every effort has been made to ensure that
the material on FederalRegister.gov is accurately displayed, consistent with
the official SGML-based PDF version on govinfo.gov, those relying on it for
legal research should verify their results against an official edition of
the Federal Register. Until the ACFR grants it official status, the XML
rendition of the daily Federal Register on FederalRegister.gov does not
provide legal notice to the public or judicial notice to the courts.
Rule
Enter a search term or FR citation e.g.
88 FR 38230 FR 78782024-13208USDA09/05/24RULE0503-AA39SORN
Choosing an item from
full text search results
will bring you to those results. Pressing enter in the search box
will also bring you to search results.
Choosing an item from
suggestions
will bring you directly to the content.
This table of contents is a navigational tool, processed from the
headings within the legal text of Federal Register documents.
This repetition of headings to form internal navigation links
has no substantive legal effect.
Comments are no longer being accepted.
See DATES for details.
Regulations.gov Data
Enhanced Content - Regulations.gov Data
FederalRegister.gov retrieves relevant information about this document
from Regulations.gov to provide users with additional context. This
information is not part of the official Federal Register document.
Medical Devices; Quality Management System Regulation Technical Amendments
Document page views are updated periodically throughout the day and are
cumulative counts for this document. Counts are subject to sampling,
reprocessing and revision (up or down) throughout the day.
Page views
7,398
as of
07/15/2026 at 6:15 pm EDT
Other Formats
Enhanced Content - Other Formats
This document is also available in the following formats:
This PDF is FR Doc. 2025-21955 as it appeared on Public Inspection on
12/03/2025 at 8:45 am.
It was viewed
160
times while on Public Inspection.
If you are using public inspection listings for legal research, you
should verify the contents of the documents against a final, official
edition of the Federal Register. Only official editions of the
Federal Register provide legal notice of publication to the public and judicial notice
to the courts under 44 U.S.C. 1503 & 1507.
Learn more here.
Published Document: 2025-21955 (90 FR 55978)
This document has been published in the Federal Register. Use the PDF linked in the document sidebar for the official electronic format.
( printed page 55978)
AGENCY:
Food and Drug Administration, HHS.
ACTION:
Final rule; technical amendments.
SUMMARY:
The Food and Drug Administration (FDA, the Agency, or we) is amending certain medical device regulations to revise references and language in existing Code of Federal Regulations (CFR) provisions to conform with the final rule “Medical Devices; Quality System Regulation Amendments” (QMSR Final Rule). This rule does not impose any new requirements on affected parties. This action is editorial in nature to correct errors, conform regulatory references, and ensure accuracy and clarity in the Agency's regulations.
DATES:
This rule is effective February 2, 2026.
FOR FURTHER INFORMATION CONTACT:
Daniel Schieffer, Office of Policy, Center for Devices and Radiological Health, 10903 New Hampshire Ave., Bldg. 66, Rm. 5562, Silver Spring, MD 20993, 301-796-3350,
Daniel.Schieffer@fda.hhs.gov.
SUPPLEMENTARY INFORMATION:
I. Background
As a part of this technical amendment, FDA is making changes to 21 CFR parts 801, 803, 812, 860, 862, 864, 866, 868, 872, 874, 876, 878, 880, 882, 886, 888, 890, and 892 to revise existing regulatory provisions for consistency with the QMSR Final Rule [1]
to correct errors, conform regulatory references, and ensure accuracy and clarity in the Agency's medical device regulations. The changes published in this notice are non-substantive and editorial in nature.
On February 2, 2024, FDA issued the QMSR Final Rule. This rule amended the device current good manufacturing practice (CGMP) requirements of the Quality System Regulation (QSR) to harmonize and modernize the regulation. The QMSR incorporates by reference an international standard for device quality management systems (ISO 13485:2016, Medical devices—Quality management systems—Requirements for regulatory purposes) and establishes additional requirements and provisions that clarify certain expectations and concepts used in ISO 13485. The purpose of this technical amendment is to update references from the QSR in existing FDA regulations to references to the QMSR that becomes effective on February 2, 2026, and to correct errors and ensure consistency and clarity in FDA's regulations.
II. Description of the Technical Amendments
We are amending 179 sections of Title 21 of the CFR, spread throughout 18 parts, in this technical amendment. Each section that is being modified includes a change to conform existing regulations for consistency with the QMSR. FDA is also clarifying the authority citations for three parts. The changes made in this technical amendment can be summarized as follows:
One hundred and sixty-two of these amendments change the wording in the classification regulations for certain class I device types.[2]
Specifically, these classification regulations currently exempt their respective devices, with various qualifications depending on the device, from the majority of the QSR. The QSR provisions that are not exempted in all of these sections (except one [3]
) are §§ 820.180 and 820.198 (21 CFR 820.180 and 820.198). This amendment updates the references to the non-exempted activities for these devices from §§ 820.180 (Records—General requirements) and 820.198 (Records—Complaint files) in the QSR to the equivalent requirement in the QMSR (§ 820.35 (Control of Records) (21 CFR 820.35)).
Section 801.30(a)(2) (21 CFR 801.30(a)(2)), which is a specific exemption to the unique device identification requirement, is being modified in the same way as the classification regulations in the preceding bullet by changing references from §§ 820.180 and 820.198 in the QSR to the equivalent requirement in the QMSR (§ 820.35).
In the classification regulation for the Keratoscope (§ 886.1350(b) (21 CFR 886.1350(b))), in addition to the updated language described in the first bullet, we are also adding the words “Classification. Class I (general controls).” This is because the statement that the device was classified as class I [4]
was inadvertently deleted from the CFR when the section was last edited in 2000.[5]
Nine class II classification regulations [6]
are being modified to change references to § 820.30 (Design Controls) (21 CFR 820.30) in the QSR to the equivalent provision in the QMSR (§ 820.10(c) (21 CFR 820.10(c))).
In part 812 (Investigational Device Exemptions) an additional two sections (§§ 812.1(a) and 812.35(a)(3)(iii)(A) (21 CFR 812.1(a) and 812.35(a)(3)(iii)(A))) are being modified to also change references from § 820.30 in the QSR to the equivalent provision in the QMSR (§ 820.10(c)). A second revision to § 812.35 revises § 812.35(a)(3)(iv)(A) to insert a clarifying reference to § 820.10(c) of the QMSR.
One change in § 801.45(e) (21 CFR 801.45(e)) changes a reference from the design history file in the QSR to the design and development files of the QMSR.
Four sections are modified in this technical amendment by updating the phrase “quality system regulation” to
( printed page 55979)
the phrase “quality management system regulation.”
FDA is clarifying the authority citations in parts 862, 876, and 878 to italicize the `l' in21 U.S.C. 360l
to more clearly distinguish it from a numeral 1.
In some of the edited sections we have also made minor wording changes to standardize language across sections and/or correct non-substantive typographic errors. The changes to all 179 sections are non-substantive and are intended to conform existing regulations to the requirements in the QMSR that mirror the requirements in the QSR. This rule does not impose any new regulatory requirements on affected parties. The amendments are editorial in nature and should not be construed as modifying any substantive standards or requirements.
III. Notice and Public Comment
Publication of this document constitutes final action under the Administrative Procedure Act (APA) (5 U.S.C. 551-559). Section 553 of the APA generally exempts “rules of agency organization, procedure, or practice” from proposed rulemaking (
i.e.,
notice and comment rulemaking) (5 U.S.C. 553(b)(4)(A)). Rules are also exempt when an agency finds “good cause” that notice and comment rulemaking procedures would be “impracticable, unnecessary, or contrary to the public interest” (5 U.S.C. 553(b)(4)(B)).
FDA has determined that this rulemaking meets the APA's notice and comment exemption requirements under 5 U.S.C. 553(b)(4)(B). All the revisions in this rule are technical or non-substantive changes. These revisions update the language in certain regulations to be consistent with other regulations and the FD&C Act without changing the actions required under the regulations. Such technical, non-substantive changes are “a routine determination, insignificant in nature and impact, and inconsequential to the industry and to the public.”
Mack Trucks, Inc.
v.
EPA,
682 F.3d 87, 94 (D.C. Cir. 2012) (quotation marks and citation omitted). FDA accordingly for good cause finds that notice and public procedure thereon are unnecessary for these amendments.
IV. Effective Date
This amendment is effective on February 2, 2026. This is the same date that the QMSR final rule becomes effective.
General exceptions from the requirement for the label of a device to bear a unique device identifier.
(a) * * *
(2) A class I device that FDA has by regulation exempted from the good manufacturing practice requirements of part 820 of this chapter, exclusive of any continuing requirement for recordkeeping under § 820.35 of this chapter.
* * * * *
3. In § 801.45, revise paragraph (e) to read as follows:
Devices that must be directly marked with a unique device identifier.
* * * * *
(e)
Exception to be noted in design and development files.
A labeler that decides to make use of an exception under paragraph (d) of this section must document the basis of that decision in the design and development files required by § 820.10(c) of this chapter.
PART 803—MEDICAL DEVICE REPORTING
4. The authority citation for part 803 continues to read as follows:
What are the requirements for establishing and maintaining MDR files or records that apply to me?
* * * * *
(e) If you are a manufacturer, you may maintain MDR event files as part of your complaint file, under part 820 of this chapter, if you prominently identify these records as MDR reportable events. We will not consider your submitted MDR report to comply with this part unless you evaluate an event in accordance with the quality management system requirements described in part 820 of this chapter. You must document and maintain in your MDR event files an explanation of why you did not submit or could not obtain any information required by this part, as well as the results of your evaluation of each event.
PART 812—INVESTIGATIONAL DEVICE EXEMPTIONS
6. The authority citation for part 812 continues to read as follows:
(a) The purpose of this part is to encourage, to the extent consistent with the protection of public health and safety and with ethical standards, the discovery and development of useful devices intended for human use, and to that end to maintain optimum freedom for scientific investigators in their pursuit of this purpose. This part provides procedures for the conduct of clinical investigations of devices. An approved investigational device exemption (IDE) permits a device that
( printed page 55980)
otherwise would be required to comply with a performance standard or to have premarket approval to be shipped lawfully for the purpose of conducting investigations of that device. An IDE approved under § 812.30 or considered approved under § 812.2(b) exempts a device from the requirements of the following sections of the Federal Food, Drug, and Cosmetic Act (the act) and the regulations in this chapter issued thereunder: Misbranding under section 502 of the act, registration, listing, and premarket notification under section 510, performance standards under section 514, premarket approval under section 515, a banned device regulation under section 516, records and reports under section 519, restricted device requirements under section 520(e), good manufacturing practice requirements under section 520(f) except for the requirements found in § 820.10(c), if applicable (unless the sponsor states an intention to comply with these requirements under § 812.20(b)(3) or § 812.140(b)(4)(v)) and color additive requirements under section 721.
* * * * *
8. In § 812.35, revise paragraphs (a)(3)(iii)(A) and (a)(3)(iv)(A) to read as follows:
(A) Credible information to support developmental changes in the device (including manufacturing changes) includes data generated under the design and development activities of § 820.10(c) of this chapter, preclinical/animal testing, peer reviewed published literature, or other reliable information such as clinical information gathered during a trial or marketing.
* * * * *
(iv) * * *
(A) For a developmental or manufacturing change to the device, the notice shall include a summary of the relevant information gathered during the course of the investigation upon which the change was based; a description of the change to the device or manufacturing process (cross-referenced to the appropriate sections of the original device description or manufacturing process); and, if the design and development activities of § 820.10(c) of this chapter were used to assess the change, a statement that no new risks were identified by appropriate risk analysis and that the verification and validation testing, as appropriate, demonstrated that the design outputs met the design input requirements. If another method of assessment was used, the notice shall include a summary of the information which served as the credible information supporting the change.
* * * * *
PART 860—MEDICAL DEVICE CLASSIFICATION PROCEDURES
9. The authority citation for part 860 continues to read as follows:
Exemptions from sections 510, 519, and 520(f) of the Federal Food, Drug, and Cosmetic Act.
(a) A panel recommendation to the Commissioner that a device be classified or reclassified into class I will include a recommendation as to whether the device should be exempted from some or all of the requirements of one or more of the following sections of the Federal Food, Drug, and Cosmetic Act: Section 510 (registration, product listing, and premarket notification), section 519 (records and reports) and section 520(f) (good manufacturing practice requirements of the quality management system regulation), and, in the case of a recommendation for classification into class II, whether the device should be exempted from the premarket notification requirement under section 510.
* * * * *
11. In § 860.84, revise paragraph (d)(4) to read as follows:
Classification procedures for “preamendments devices.”
* * * * *
(d) * * *
(4) In the case of a recommendation for classification into class I, a recommendation as to whether the device should be exempted from the requirements of one or more of the following sections of the Federal Food, Drug, and Cosmetic Act: Section 510 (registration, product listing, and premarket notification), section 519 (records and reports), and section 520(f) (good manufacturing practice requirements of the quality management system regulation) and, in the case of a recommendation for classification into class II, whether the device should be exempted from the premarket notification requirement under section 510, in accordance with § 860.15;
* * * * *
12. In § 860.240, revise paragraph (c)(2) to read as follows:
(2) That implementation of Quality Management System Regulation (part 820 of this chapter) requirements, in addition to other general controls and any specified special controls, provide adequate assurance that critical and/or novel manufacturing processes produce devices that meet specifications necessary to ensure reasonable assurance of safety and effectiveness.
PART 862—CLINICAL CHEMISTRY AND CLINICAL TOXICOLOGY DEVICES
13. The authority citation for part 862 is revised to read as follows:
(2) As part of the risk management activities performed as part of your 21 CFR 820.10(c) design and development activities, you must document the appropriate end user device training program provided in your premarket notification submission to satisfy the special control in paragraph (b)(1) of this section that will be offered while marketing the device as part of your efforts to mitigate the risk of incorrect interpretation of test results.
* * * * *
15. In § 862.2050, revise paragraph (b) to read as follows:
General purpose laboratory equipment labeled or promoted for a specific medical use.
* * * * *
(b)
Classification.
Class I (general controls). The device is identified in paragraph (a) of this section and is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9. The device is also exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
16. In § 862.2270, revise paragraph (b) to read as follows:
Thin-layer chromatography system for clinical use.
* * * * *
( printed page 55981)
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 862.9. Particular components of TLC systems,
i.e.,
the thin-layer chromatography apparatus, TLC atomizer, TLC developing tanks, and TLC ultraviolet light, are exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
PART 864—HEMATOLOGY AND PATHOLOGY DEVICES
17. The authority citation for part 864 continues to read as follows:
(b)
Classification.
Class I (general controls). These devices are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 864.9. These devices are also exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
19. In § 864.2240, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). These devices are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 864.9. If the devices are not labeled or otherwise represented as sterile, they are exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
20. In § 864.3010, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). These devices are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 864.9. The devices are also exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
21. In § 864.3250, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 864.9. If the device is not labeled or otherwise represented as sterile, it is exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
22. In § 864.3260, revise paragraph (b) to read as follows:
OTC test sample collection systems for drugs of abuse testing.
* * * * *
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification requirements in part 807, subpart E of this chapter subject to the limitations in § 864.9 if it is sold, distributed, and used in accordance with the restrictions set forth in § 809.40 of this chapter. If the device is not labeled or otherwise represented as sterile, it is exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning complaint files under § 820.35(a) of this chapter.
23. In § 864.3600, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). These devices are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 864.9. If the device is not labeled or otherwise represented as sterile, it is exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
24. In § 864.4010, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 864.9. If the device is not labeled or otherwise represented as sterile, it is exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
25. In § 864.7010, revise paragraph (b)(3) to read as follows:
Flow cytometric test system for hematopoietic neoplasms.
* * * * *
(b) * * *
(3) As part of the risk management activities performed under 21 CFR 820.10(c) design and development, product labeling and instruction manuals must include clear examples of all expected phenotypic patterns and gating strategies using well-defined clinical samples representative of both abnormal and normal cellular populations. These samples must be selected based upon the indications described in paragraph (b)(1)(i) of this section.
PART 866—IMMUNOLOGY AND MICROBIOLOGY DEVICES
26. The authority citation for part 866 continues to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 866.9. The device is also exempt from the good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
28. In § 866.2180, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 866.9. The device is also exempt from the good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
29. In § 866.2190, revise paragraph (b)(7) to read as follows:
Automated image assessment system for microbial colonies on solid culture media.
* * * * *
(b) * * *
(7) Under 21 CFR 820.10(c) design and development, device manufacturers must, as appropriate:
(i) Conduct human factors/usability validation testing with the final version of the labeling and related materials to adequately mitigate the risk of failure to operate the instrument correctly.
(ii) Document a device training program that will be offered to the end user to adequately mitigate the risk of failure to operate the instrument correctly.
30. In § 866.2440, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 866.9. The device is also exempt from the good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
31. In § 866.2540, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 866.9. The device is also exempt from the good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
32. In § 866.2600, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 866.9. The device is also exempt from the good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
33. In § 866.3215, revise paragraph (b)(6) to read as follows:
Device to detect and measure non-microbial analyte(s) in human clinical specimens to aid in assessment of patients with suspected sepsis.
* * * * *
(b) * * *
(6) As part of the risk management activities performed under 21 CFR 820.10(c) design and development, you must document an appropriate end user device training program that will be offered as part of your efforts to mitigate the risk of failure to correctly operate the instrument.
* * * * *
34. In § 866.3309, revise paragraph (b)(7) to read as follows:
Herpes virus nucleic acid-based cutaneous and mucocutaneous lesion panel.
* * * * *
(b) * * *
(7) The risk management activities performed as part of the manufacturer's 21 CFR 820.10(c) design and development activities must document an appropriate end user device training program that will be offered as part of efforts to mitigate the risk of failure to correctly operate the instrument.
35. In § 866.3361, revise paragraph (b)(4) to read as follows:
Mass spectrometer system for clinical use for the identification of microorganisms.
* * * * *
(b) * * *
(4) As part of the risk management activities performed under 21 CFR 820.10(c) design and development, you must document an appropriate end user device training program that will be offered as part of your efforts to mitigate the risk of failure to correctly operate the instrument.
* * * * *
36. In § 866.3960, revise paragraph (b)(8) to read as follows:
Nucleic acid-based device for the amplification, detection, and identification of microbial pathogens directly from whole blood specimens.
* * * * *
(b) * * *
(8) As part of the risk management activities performed under 21 CFR 820.10(c) design and development, you must document an appropriate end user device training program that will be offered as part of your efforts to mitigate the risk of failure to correctly operate the instrument.
37. In § 866.3970, revise paragraph (b)(10) to read as follows:
Device to detect and identify microbial pathogen nucleic acids in cerebrospinal fluid.
* * * * *
(b) * * *
(10) As part of the risk management activities performed under 21 CFR 820.10(c) design and development, you must document an appropriate end user device training program that will be offered as part of your efforts to mitigate the risk of failure to correctly operate the instrument.
PART 868—ANESTHESIOLOGY DEVICES
38. The authority citation for part 868 continues to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 868.9. The device is also exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
40. In § 868.1965, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 868.9. The device is also exempt from the current
( printed page 55983)
good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
41. In § 868.5220, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 868.9. If the device is not labeled or otherwise represented as sterile, it is exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
42. In § 868.5420, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 868.9. If the device is not labeled or otherwise represented as sterile, it is exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
43. In § 868.5760, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 868.9. If the device is not labeled or otherwise represented as sterile, it is exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
44. In § 868.5795, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 868.9. If the device is not labeled or otherwise represented as sterile, it is exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
45. In § 868.6175, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 868.9. The device is also exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
46. In § 868.6225, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 868.9. The device is also exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
PART 872—DENTAL DEVICES
47. The authority citation for part 872 continues to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 872.9. If the device is not labeled or otherwise represented as sterile, it is also exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
49. In § 872.3140, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 872.9. If the device is not labeled or otherwise represented as sterile, the device is exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
50. In § 872.3150, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 872.9. If the device is not labeled or otherwise represented as sterile, the device is exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
51. In § 872.3220, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 872.9. If the device is not labeled or otherwise represented as sterile, the device is exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
52. In § 872.3670, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 872.9. If
( printed page 55984)
the device is not labeled or otherwise represented as sterile, it is exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
53. In § 872.3730, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 872.9. If the device is not labeled or otherwise represented as sterile, it is exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
54. In § 872.6010, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 872.9. If the device is not labeled or otherwise represented as sterile, it is exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
55. In § 872.6050, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 872.9. If the device is not labeled or otherwise represented as sterile, it is exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
56. In § 872.6140, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 872.9. If the device is not labeled or otherwise represented as sterile, it is exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
57. In § 872.6200, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 872.9. If the device is not labeled or otherwise represented as sterile, it is exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
58. In § 872.6290, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 872.9. If the device is not labeled or otherwise represented as sterile, it is exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
59. In § 872.6300, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 872.9. If the device is not labeled or otherwise represented as sterile, it is exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
60. In § 872.6570, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 872.9. If the device is not labeled or otherwise represented as sterile, it is exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
61. In § 872.6650, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 872.9. If the device is not labeled or otherwise represented as sterile, it is exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
62. In § 872.6670, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 872.9. If the device is not labeled or otherwise represented as sterile, it is exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
63. In § 872.6855, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 872.9. If the device is not labeled or otherwise represented as sterile, it is exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements
( printed page 55985)
concerning records and complaint files under § 820.35 of this chapter.
64. In § 872.6870, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 872.9. If the device is not labeled or otherwise represented as sterile, it is exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
65. In § 872.6880, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 872.9. If the device is not labeled or otherwise represented as sterile, it is exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
66. In § 872.6890, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 872.9. If the device is not labeled or otherwise represented as sterile, it is exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
PART 874—EAR, NOSE, AND THROAT DEVICES
67. The authority citation for part 874 continues to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 874.9. If the device is not labeled or otherwise represented as sterile, it is exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
69. In § 874.3540, revise paragraph (b) to read as follows:
Prosthesis modification instrument for ossicular replacement surgery.
* * * * *
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 874.9. If the device is not labeled or otherwise represented as sterile, it is exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
70. In § 874.5220, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 874.9. If the device is not labeled or otherwise represented as sterile, it is exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
PART 876—GASTROENTEROLOGY-UROLOGY DEVICES
71. The authority citation for part 876 is revised to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9. The device is exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
73. In § 876.5250, revise paragraph (b)(2) to read as follows:
(2)
Class I (general controls).
For a urine collector and accessories not intended to be connected to an indwelling catheter, the device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9. If the device is not labeled or otherwise represented as sterile, it is exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
74. In § 876.5920, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9. The device is also exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
75. In § 876.5970, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9. The device is also exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
( printed page 55986)
PART 878—GENERAL AND PLASTIC SURGERY DEVICES
76. The authority citation for part 878 is revised to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 878.9. If the device is intended for use without an external prosthesis adhesive to fasten it to the body, the device is exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
78. In § 878.3910, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9. If the device is not labeled or otherwise represented as sterile, it is exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
PART 880—GENERAL HOSPITAL AND PERSONAL USE DEVICES
79. The authority citation for part 880 continues to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 880.9. The device also is exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
81. In § 880.2740, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 880.9. The device also is exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
82. In § 880.2930, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9. The device is also exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
83. In § 880.5075, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 880.9. The device also is exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
84. In § 880.5120, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 880.9. The device is also exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
85. In § 880.5150, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 880.9. The device also is exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
86. In § 880.5160, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 880.9. If the device is not labeled or otherwise represented as sterile, it is also exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
87. In § 880.5270, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 880.9. If the device is not labeled or otherwise represented as sterile, it is exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
88. In § 880.5300, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 880.9. If the device is not labeled or otherwise represented as sterile, it is also exempt
( printed page 55987)
from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
89. In § 880.5640, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 880.9. If the device is not labeled or otherwise represented as sterile, it is also exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
90. In § 880.5680, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). Except when the device is an infant positioner for prescription use in highly monitored settings or an infant sleep position holder, it is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9. The device is exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
91. In § 880.5780, revise paragraph (b)(2) to read as follows:
(2)
Classification.
Class I. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 880.9. The device is also exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
92. In § 880.5820, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 880.9. The device also is exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
93. In § 880.6025, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 880.9. If the device is not labeled or otherwise represented as sterile, it is also exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
94. In § 880.6050, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 880.9. If the device is not labeled or otherwise represented as sterile, it is also exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
95. In § 880.6060, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 880.9. If the device is not labeled or otherwise represented as sterile, it is also exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
96. In § 880.6070, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 880.9. The device is also exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
97. In § 880.6080, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 880.9. The device is also exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
98. In § 880.6085, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 880.9. The device is also exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
99. In § 880.6140, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 880.9. The device is also exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
( printed page 55988)
100. In § 880.6185, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 880.9. If the device is not labeled or otherwise represented as sterile, it is also exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
101. In § 880.6190, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 880.9. If the device is not labeled or otherwise represented as sterile, it is also exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
102. In § 880.6200, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 880.9. The device also is exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
103. In § 880.6230, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 880.9. If the device is not labeled or otherwise represented as sterile, it is also exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
104. In § 880.6265, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 880.9. If the device is not labeled or otherwise represented as sterile, it is also exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
105. In § 880.6350, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 880.9. The device also is exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
106. In § 880.6430, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 880.9. The device is also exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
107. In § 880.6450, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 880.9. The device is also exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
108. In § 880.6730, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 880.9. The device also is exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
109. In § 880.6785, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 880.9. The device is also exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
110. In § 880.6800, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 880.9. The device also is exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
111. In § 880.6900, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter,
( printed page 55989)
subject to the limitations in § 880.9. The device is also exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
112. In § 880.6960, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 880.9. If the device is not labeled or otherwise represented as sterile, it is also exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
113. In § 880.6980, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 880.9. If the device is not labeled or otherwise represented as sterile, it is also exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
PART 882—NEUROLOGICAL DEVICES
114. The authority citation for part 882 continues to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 882.9. The device is also exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
116. In § 882.1500, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 882.9. The device is also exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
117. In § 882.1525, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 882.9. The device is also exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
118. In § 882.1700, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 882.9. The device is also exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
PART 886—OPHTHALMIC DEVICES
119. The authority citation for part 886 continues to read as follows:
(b)
Classification.
Class I. The AC-powered device and the manual device are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 886.9. The manual device is also exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
121. In § 886.1150, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 886.9. The device is also exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
122. In § 886.1170, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 886.9. The device is also exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
123. In § 886.1190, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 886.9. The device is also exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
124. In § 886.1200, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter,
( printed page 55990)
subject to the limitations in § 886.9. The device is also exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
125. In § 886.1320, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 886.9. The device is also exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
126. In § 886.1330, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 886.9. The device is also exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
127. In § 886.1350, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 886.9. The battery-powered device is exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
128. In § 886.1375, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 886.9. The device is also exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
129. In § 886.1380, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 886.9. The device is also exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
130. In § 886.1390, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 886.9. The device is also exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
131. In § 886.1395, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 886.9. The device is also exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
132. In § 886.1400, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 886.9. The device is also exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
133. In § 886.1415, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 886.9. The device is also exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
134. In § 886.1460, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 886.9. The device is also exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
135. In § 886.1500, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 886.9. The device is also exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
136. In § 886.1605, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 886.9. The
( printed page 55991)
device is also exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
137. In § 886.1650, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 886.9. The device is also exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
138. In § 886.1655, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 886.9. The device is also exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
139. In § 886.1665, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 886.9. The device is also exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
140. In § 886.1700, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The AC-powered device and the manual device are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 886.9. The manual device is also exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
141. In § 886.1770, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 886.9. The device is also exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
142. In § 886.1780, revise paragraph (b)(2) to read as follows:
(2) Class I (general controls) for the battery-powered device. The class I battery-powered device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 886.9. The battery-powered device is exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
143. In § 886.1790, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 886.9. The device is also exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
144. In § 886.1810, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The AC-powered device and the battery-powered device are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 886.9. The battery-powered device is also exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
145. In § 886.1840, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 886.9. The device is also exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
146. In § 886.1860, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The AC-powered device and the battery-powered device are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 886.9. The battery-powered device is also exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
147. In § 886.1870, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The AC-powered device and the battery-powered device are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 886.9. The battery-powered device is also exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
148. In § 886.1880, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 886.9. The device is also exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
149. In § 886.1905, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 886.9. The device is also exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
150. In § 886.1910, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The AC-powered device and the battery-powered device are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 886.9. The battery-powered device is also exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
151. In § 886.4230, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 886.9. The device is also exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
152. In § 886.4445, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 886.9. The device is also exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
153. In § 886.4750, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). When made only of plastic or aluminum, the device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 886.9. When made only of plastic or aluminum, the devices are exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
154. In § 886.4770, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 886.9. The device is also exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
155. In § 886.4855, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The AC-powered device and the manual device are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 886.9. The manual device is also exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
156. In § 886.5120, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 886.9. The device is also exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
157. In § 886.5540, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 886.9. The device is also exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
158. In § 886.5600, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 886.9. The device is also exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
159. In § 886.5800, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 886.9. The device is also exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements
( printed page 55993)
concerning records and complaint files under § 820.35 of this chapter.
160. In § 886.5810, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 886.9. The device is also exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
161. In § 886.5870, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 886.9. The device is also exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
162. In § 886.5910, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 886.9. The device is also exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
163. In § 886.5915, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The AC-powered device and the battery-powered device are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 886.9. The battery-powered device is also exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
PART 888—ORTHOPEDIC DEVICES
164. The authority citation for part 888 continues to read as follows:
Non-powered orthopedic traction apparatus and accessories.
* * * * *
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 888.9. The device is also exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
166. In § 888.5890, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 888.9. The device is also exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
167. In § 888.5940, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 888.9. The device is also exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
168. In § 888.5980, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 888.9. The device is also exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
PART 890—PHYSICAL MEDICINE DEVICES
169. The authority citation for part 890 continues to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 890.9. The device is also exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
171. In § 890.3075, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 890.9. The device is also exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
172. In § 890.3150, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 890.9. The device is also exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this
( printed page 55994)
chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
173. In § 890.3410, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 890.9. The device is also exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
174. In § 890.3420, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 890.9. The device is also exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
175. In § 890.3475, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 890.9. The device is also exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
176. In § 890.3490, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 890.9. The device is also exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
177. In § 890.3520, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 890.9. The device is also exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
178. In § 890.3640, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 890.9. The device is also exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
179. In § 890.3665, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 890.9. The device is also exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
180. In § 890.3675, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 890.9. The device is also exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
181. In § 890.3700, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 890.9. The device is also exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
182. In § 890.3790, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 890.9. The device is also exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
183. In § 890.3825, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 890.9. The device is also exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
184. In § 890.3910, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). If the device is not intended for use as a protective restraint as defined in § 880.6760 of this chapter, it is exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the
( printed page 55995)
limitations in § 890.9. The device is also exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
185. In § 890.3940, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 890.9. The device is also exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
186. In § 890.5050, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 890.9. If the device is not labeled or otherwise represented as sterile, the device is also exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
187. In § 890.5125, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 890.9. The device is also exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
188. In § 890.5350, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 890.9. The device is also exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
189. In § 890.5370, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 890.9. The device is also exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
190. In § 890.5700, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807. The device also is exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820, except for requirements concerning records and complaint files under § 820.35 of this chapter.
191. In § 890.5730, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 890.9. The device is also exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
192. In § 890.5925, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 890.9. The device is also exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
PART 892—RADIOLOGY DEVICES
193. The authority citation for part 892 continues to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 892.9. The device is also exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
195. In § 892.1940, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 892.9. The device is also exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
196. In § 892.1950, revise paragraph (b) to read as follows:
(b)
Classification.
Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 892.9. The device is also exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
( printed page 55996)
197. In § 892.5725, revise paragraph (b)(2) to read as follows:
(2) The risk management activities performed as part of the manufacturer's § 820.10(c) of this chapter design and development activities must document an appropriate end user initial training program which will be offered as part of efforts to mitigate the risk of failure to correctly operate the device, including, but not limited to, documentation of an appropriate end user initial training program on the proper spacer deployment technique.
* * * * *
Lowell M. Zeta,
Acting Deputy Commissioner for Policy, Legislation, and International Affairs.
Footnotes
1.
In this technical amendment, FDA uses the terms below in the following manner: when referring to 21 CFR part 820 as amended and effective on February 2, 2026, FDA uses the terms “Quality Management System Regulation” or “QMSR.” When referring to the regulation at 21 CFR part 820 in effect before February 2, 2026, FDA uses the terms “Quality System Regulation” or “QSR.”
2.
Class I devices are those devices for which the general controls within the Federal Food, Drug, and Cosmetic Act (FD&C Act) (controls authorized by or under section 501, 502, 510, 516, 518, 519, or 520 (21 U.S.C. 351, 352, 360, 360f, 360h, 360i, or 360j) or any combination of such sections) are sufficient to provide reasonable assurance of safety and effectiveness; or those devices for which insufficient information exists to determine that general controls are sufficient to provide reasonable assurance of safety and effectiveness or to establish special controls to provide such assurance, but because the devices are not purported or represented to be for a use in supporting or sustaining human life or for a use which is of substantial importance in preventing impairment of human health, and do not present a potential unreasonable risk of illness or injury, are to be regulated by general controls (section 513(a)(1)(A) of the FD&C Act (21 U.S.C. 360c(a)(1)(A))).
3.
Section 864.3260(b) (21 CFR 864.3260(b)) currently exempts over-the-counter (OTC) test sample collection systems for drugs of abuse testing, if the device is not labeled or otherwise represented as sterile, from all of the QSR except for § 820.198.
6.
Class II devices are those devices for which general controls by themselves are insufficient to provide reasonable assurance of safety and effectiveness, and for which there is sufficient information to establish special controls to provide such assurance, including the issuance of performance standards, postmarket surveillance, patient registries, development and dissemination of guidelines, recommendations, and other appropriate actions the Agency deems necessary to provide such assurance (section 513(a)(1)(B) of the FD&C Act).