This site displays a prototype of a “Web 2.0” version of the daily
Federal Register. It is not an official legal edition of the Federal
Register, and does not replace the official print version or the official
electronic version on GPO’s govinfo.gov.
The documents posted on this site are XML renditions of published Federal
Register documents. Each document posted on the site includes a link to the
corresponding official PDF file on govinfo.gov. This prototype edition of the
daily Federal Register on FederalRegister.gov will remain an unofficial
informational resource until the Administrative Committee of the Federal
Register (ACFR) issues a regulation granting it official legal status.
For complete information about, and access to, our official publications
and services, go to
About the Federal Register
on NARA's archives.gov.
The OFR/GPO partnership is committed to presenting accurate and reliable
regulatory information on FederalRegister.gov with the objective of
establishing the XML-based Federal Register as an ACFR-sanctioned
publication in the future. While every effort has been made to ensure that
the material on FederalRegister.gov is accurately displayed, consistent with
the official SGML-based PDF version on govinfo.gov, those relying on it for
legal research should verify their results against an official edition of
the Federal Register. Until the ACFR grants it official status, the XML
rendition of the daily Federal Register on FederalRegister.gov does not
provide legal notice to the public or judicial notice to the courts.
Rule
Enter a search term or FR citation e.g.
88 FR 38230 FR 78782024-13208USDA09/05/24RULE0503-AA39SORN
Choosing an item from
full text search results
will bring you to those results. Pressing enter in the search box
will also bring you to search results.
Choosing an item from
suggestions
will bring you directly to the content.
Medical Devices; Exemption From Premarket Notification: Radiology Computer-Aided Detection and/or Diagnosis Devices and Computer-Aided Triage and Notification Devices
This table of contents is a navigational tool, processed from the
headings within the legal text of Federal Register documents.
This repetition of headings to form internal navigation links
has no substantive legal effect.
Medical Devices; Exemption From Premarket Notification: Radiology Computer-Aided Detection and/or Diagnosis Devices and Computer-Aided Triage and Notification Devices
Comments are no longer being accepted.
See DATES for details.
Regulations.gov Data
Enhanced Content - Regulations.gov Data
FederalRegister.gov retrieves relevant information about this document
from Regulations.gov to provide users with additional context. This
information is not part of the official Federal Register document.
Requests that the FDA create an alternative pathway for radiology Computer-Aided Detection and/or Diagnosis (CAD) and Computer-Aided Triage and Notification (CADt) devices, partially exempting these devices from the premarket notification requirements of section 510(k) of the FD&C Act.
Document page views are updated periodically throughout the day and are
cumulative counts for this document. Counts are subject to sampling,
reprocessing and revision (up or down) throughout the day.
Page views
1,879
as of
11/18/2026 at 10:15 pm EST
Other Formats
Enhanced Content - Other Formats
This document is also available in the following formats:
This PDF is FR Doc. 2026-19074 as it appeared on Public Inspection on
09/16/2026 at 8:45 am.
It was viewed
130
times while on Public Inspection.
If you are using public inspection listings for legal research, you
should verify the contents of the documents against a final, official
edition of the Federal Register. Only official editions of the
Federal Register provide legal notice of publication to the public and judicial notice
to the courts under 44 U.S.C. 1503 & 1507.
Learn more here.
Published Document: 2026-19074 (91 FR 58817)
This document has been published in the Federal Register. Use the PDF linked in the document sidebar for the official electronic format.
AGENCY:
Food and Drug Administration, HHS.
ACTION:
Final order.
SUMMARY:
The Food and Drug Administration (FDA) is publishing an order setting forth its final determination regarding a partial exemption from the premarket notification requirements for radiology computer-aided detection and/or diagnosis devices and computer-aided triage and notification devices that was the subject of a notice published in the
Federal Register
of December 29, 2025. That notice announced FDA's receipt of a petition that requested exemption from the premarket notification requirements for the following generic device types when certain conditions described in the petition were met: radiological computer-assisted diagnostic software for lesions suspicious of cancer; medical image analyzers; radiological computer aided triage and notification software; and radiological computer-assisted detection and diagnosis software. FDA denied the petition in a response issued to the petitioner on April 1, 2026. FDA is publishing this order in accordance with procedures established by the Federal Food, Drug, and Cosmetic Act (FD&C Act).
DATES:
This order is effective September 17, 2026.
FOR FURTHER INFORMATION CONTACT:
Gugandeep Kaur, Center for Devices and Radiological Health, Food and Drug Administration, 10903 New Hampshire Ave., Bldg. 66, Rm. 5504, Silver Spring, MD 20993-0002, 240-402-9534.
SUPPLEMENTARY INFORMATION:
I. Background—Regulatory Authorities
The FD&C Act, as amended, establishes a comprehensive system for the regulation of medical devices intended for human use. Section 513 of the FD&C Act (21 U.S.C. 360c) establishes three classes of devices, reflecting the regulatory controls needed to provide reasonable assurance of their safety and effectiveness. The three classes of devices are class I (general controls), class II (special controls), and class III (premarket approval).
Section 513(a)(1) of the FD&C Act defines the three classes of devices. Class I devices are those devices for which the general controls of the FD&C Act (controls authorized by or under section 501, 502, 510, 516, 518, 519, or 520 (21 U.S.C. 351, 352, 360, 360f, 360h, 360i, or 360j) or any combination of such sections) are sufficient to provide reasonable assurance of safety and effectiveness of the device; or those devices for which insufficient information exists to determine that general controls are sufficient to provide reasonable assurance of safety and effectiveness or to establish special controls to provide such assurance, but because the devices are not purported or represented to be for a use in supporting or sustaining human life or for a use which is of substantial importance in preventing impairment of human health, and do not present a potential unreasonable risk of illness or injury, are to be regulated by general controls (section 513(a)(1)(A) of the FD&C Act).
Class II devices are those devices for which general controls by themselves are insufficient to provide reasonable assurance of safety and effectiveness, but for which there is sufficient information to establish special controls to provide such assurance, including the issuance of performance standards, post-market surveillance, patient registries, development and dissemination of guidelines, recommendations, and other appropriate actions FDA (the Agency or we) deems necessary to provide such assurance (section 513(a)(1)(B) of the FD&C Act).
( printed page 58818)
Class III devices are those devices for which insufficient information exists to determine that general controls and special controls would provide a reasonable assurance of safety and effectiveness, and are purported or represented to be for a use in supporting or sustaining human life or for a use which is of substantial importance in preventing impairment of human health, or present a potential unreasonable risk of illness or injury (section 513(a)(1)(C) of the FD&C Act).
Under section 510(k) of the FD&C Act and FDA's implementing regulations in part 807, subpart E (21 CFR part 807, subpart E), persons who are required to register and who propose to begin the introduction or delivery for introduction into interstate commerce for commercial distribution of a device intended for human use are required to submit a premarket notification (510(k)) to FDA. The device may not be marketed until FDA finds it “substantially equivalent” within the meaning of section 513(i) of the FD&C Act to a legally marketed device that does not require premarket approval. A premarket notification is not required for devices in certain situations, such as when they have been exempted from that requirement under section 510(m) of the FD&C Act.
The 21st Century Cures Act (Pub. L. 114-255) (Cures Act) was signed into law on December 13, 2016. Section 3054 of the Cures Act amended section 510(m) of the FD&C Act. As amended, section 510(m)(1) of the FD&C Act requires that within 90 days of the date of enactment of the Cures Act, and at least once every 5 years thereafter (as FDA determines appropriate), FDA publish in the
Federal Register
a notice containing a list of each type of class II device that FDA determines no longer requires a report under section 510(k) of the FD&C Act to provide reasonable assurance of safety and effectiveness. After providing at least a 60-day public comment period, FDA must then publish in the
Federal Register
a list representing the final determination with respect to the devices contained in the list under section 510(m)(1)(B). Additionally, section 510(m)(2) of the FD&C Act provides that FDA may exempt a class II device from the requirement to submit a report under section 510(k) of the FD&C Act, upon its own initiative or a petition of an interested person, if FDA determines that a report under section 510(k) is not necessary to assure the safety and effectiveness of the device. FDA must publish in the
Federal Register
a notice of its intent to exempt the device, or of the petition, and provide a 60-calendar-day period for public comment. If FDA fails to respond to a petition under this section within 180 days of receiving it, the petition shall be deemed granted. In addition, within 120 days after the issuance of the notice, FDA must publish an order in the
Federal Register
that sets forth its final determination regarding the exemption of the device that was the subject of the notice.
Accordingly, FDA generally considers the following factors to determine whether a report under section 510(k) is necessary or if an exemption would be appropriate for class II devices: (1) The device does not have a significant history of false or misleading claims or of risks associated with inherent characteristics of the device; (2) characteristics of the device necessary for its safe and effective performance are well established; (3) changes in the device that could affect safety and effectiveness will either (a) be readily detectable by users by visual examination or other means such as routine testing, before causing harm, or (b) not materially increase the risk of injury, incorrect diagnosis, or ineffective treatment; and (4) any changes to the device would not be likely to result in a change in the device's classification.
FDA may also consider that, even when exempting devices from the 510(k) requirements, these devices would still be subject to general limitations of exemptions. Specifically, even if a device is exempted from premarket notification requirements, a manufacturer of the device will still be required to submit a premarket notification to FDA before introducing a device or delivering it for introduction into interstate commerce for commercial distribution when the device exceeds any of the limitations of exemptions described in 21 CFR parts 862-892, in the section of each part entitled “Limitations of exemptions from section 510(k) of the Federal Food, Drug, and Cosmetic Act (the act)” (see,
e.g.,21 CFR 892.9).
In addition to the general limitations, partial limitations may limit an exemption from premarket notification requirements to specific devices within a device type when the Agency determines that the factors described in the
Federal Register
notice (63 FR 3142) and Class II 510(k) Exemption Guidance do not weigh in favor of exemption for all devices within a generic type of device. Where partial limitations exist, FDA has determined that premarket notification is necessary to provide a reasonable assurance of safety and effectiveness for devices that fall outside of the limitations.
III. Petition
On October 22, 2025, FDA received a petition requesting partial exemption from 510(k) requirements for “radiology Computer-Aided Detection and/or Diagnosis (CAD) and Computer-Aided Triage and Notification (CADt) devices” (the Subject CAD and CADt Devices) from Nancy Stade, J.D., of Rubrum Advising, LLC, 404 Pembroke Rd., Bala Cynwyd, PA 19004, on behalf of Harrison.ai (see Docket No. FDA-2025-P-5560). Specifically, the petition sought partial exemption for the following devices:
Radiological computer-assisted diagnostic software for lesions suspicious of cancer, classified under § 892.2060 (21 CFR 892.2060), product code POK (CADx).
Medical image analyzer, classified under § 892.2070 (21 CFR 892.2070), product code MYN (CADe).
Radiological computer aided triage and notification software, classified under § 892.2080 (21 CFR 892.2080), product codes QAS and QFM (CADt).
Radiological computer-assisted detection and diagnosis software, classified under § 892.2090 (21 CFR 892.2090), product codes QBS and QDQ (CADe/x).
All these devices currently require premarket notification under section 510(k) of the FD&C Act.
The petition requested exemption from the premarket notification requirements for the Subject CAD and CADt Devices when:
The manufacturer had previously obtained a 510(k);
For devices under § 892.2080 (the Subject CADt Devices), the manufacturer had at least one clearance under the same classification regulation;
For devices under §§ 892.2060, 892.2070, or 892.2090 (the Subject CAD Devices), the manufacturer had at least
( printed page 58819)
one clearance under any of those same three classification regulations;
The manufacturer implemented “a robust post-market plan, transparency, and training measures” as described in the petition; and
All existing “special controls, quality systems, establishment registration, and device listing requirements” remained in force.
On December 29, 2025, FDA published a notice of the petition in the
Federal Register
(90 FR 60730) and requested comments on it, in accordance with section 510(m)(2) of the FD&C Act. The comment period closed on February 27, 2026.
FDA considered the information available to the Agency, including comments from the public docket for the petition and
Federal Register
notice, and determined not to exempt devices classified under §§ 892.2060, 892.2070, 892.2080, and 892.2090 from the 510(k) requirements subject to the partial limitations of exemption proposed in the petition.[1]
Accordingly, FDA responded to the petition by letter dated April 1, 2026, denying the petition within the 180-day timeframe under section 510(m)(2) of the FD&C Act.
IV. Order
As discussed in the petition response issued to the petitioner on April 1, 2026, based on FDA's review of the petition's proposed partial exemption, and in consideration of the comments submitted to the docket and FDA's own assessment of each of the four factors for exemption described in the
Federal Register
notice (63 FR 3142) and the Class II 510(k) Exemption Guidance, FDA determined that the information presented in the petition does not demonstrate that premarket notification is not necessary to assure the safety and effectiveness of the Subject CAD and CADt Devices that may be covered by the petition's proposed partial exemption. Therefore, FDA denied the petition request for partial exemption from premarket notification requirements for the Subject CAD and CADt Devices and is issuing this order setting forth the final determination. FDA's response to the petition can be found in Docket No. FDA-2025-P-5560, available at
https://www.regulations.gov/docket/FDA-2025-P-5560.
Manufacturers of the Subject CAD and CADt Devices must continue to submit and receive FDA clearance of a 510(k) submission before marketing their devices, as well as comply with all other applicable requirements under the FD&C Act.
Although FDA determined that the proposal in the petition did not support a partial exemption from 510(k) requirements under section 510(m)(2) of the FD&C Act, the Agency has a longstanding commitment to develop and apply innovative approaches to the regulation of medical device software and other digital health devices to ensure their safety and effectiveness consistent with least burdensome principles. FDA supports the continued consideration of innovative and least burdensome approaches that may accelerate the availability of safe and effective devices.
V. Analysis of Environmental Impact
We have determined under 21 CFR 25.34(b) that this action is of a type that does not normally have a significant effect on the human environment. Therefore, neither an environmental assessment nor an environmental impact statement is required.
VI. Paperwork Reduction Act of 1995
This final administrative order refers to previously approved collections of information found in FDA regulations. The previously approved collections of information are subject to review by the Office of Management and Budget (OMB) under the Paperwork Reduction Act of 1995 (44 U.S.C. 3501-3521). The collections of information in 21 CFR part 820 (Quality Management System Regulation) have been approved under OMB control number 0910-0073; the collections of information in 21 CFR part 812 (Investigational Device Exemptions) have been approved under OMB control number 0910-0078; the collections of information in part 807, subpart E (Premarket Notification Procedures), have been approved under OMB control number 0910-0120; the collections of information in 21 CFR part 822 (Postmarket Surveillance) have been approved under OMB control number 0910-0449; and the collections of information under 21 CFR part 801 (Device Labeling) have been approved under OMB control number 0910-0485.
VII. References
The following reference is on display at the Dockets Management Staff (HFA-305), Food and Drug Administration, 5630 Fishers Lane, Rm. 1061, Rockville, MD 20852, 240-402-7500, and is available for viewing by interested persons between 9 a.m. and 4 p.m., Monday through Friday; it is also available electronically at
https://www.regulations.gov.
Although FDA verified the website addresses in this document, please note that websites are subject to change over time.
1. FDA Guidance, “Procedures for Class II Device Exemptions from Premarket Notification, Guidance for Industry and CDRH Staff,” February 19, 1998, available at
https://www.fda.gov/media/72685/download.
Grace R. Graham,
Deputy Commissioner for Policy, Legislation, and International Affairs.
Footnotes
1.
In considering the petition and comments submitted to the docket, FDA has assumed that all elements of petitioner's proposal were partial limitations on exemption that FDA may consider under section 510(m)(2) of the FD&C Act and we refer to them as partial limitations on exemption throughout this order.