Medical Devices; Reprocessed Single-Use Devices; Requirement for Submission of Validation Data

Document Details

Information about this document as published in the Federal Register.

Published Document

This document has been published in the Federal Register. Use the PDF linked in the document sidebar for the official electronic format.

Start Preamble

AGENCY:

Food and Drug Administration, HHS.

ACTION:

Direct final rule.

SUMMARY:

The Food and Drug Administration (FDA or we) is amending certain classification regulations for reprocessed single-use devices (SUDs) whose exemption from premarket notification (510(k)) requirements have been terminated and other reprocessed SUDs already subject to premarket notification for which validation data, as specified under the Medical Device User Fee and Modernization Act of 2002 (MDUFMA), are necessary in a 510(k). Elsewhere in this issue of the Federal Register, we are publishing a companion proposed rule, under FDA's usual procedures for notice and comment, to provide a procedural framework to finalize the rule in the event we receive any significant adverse comment and withdraw the direct final rule. This action codifies actions taken in previous Federal Register notices in accordance with MDUFMA.

DATES:

This rule is effective February 7, 2007. Submit written or electronic comments by December 11, 2006. If we receive no significant adverse comments within the specified comment period, we intend to publish a document confirming the effective date of the final rule in the Federal Register within 30 days after the comment period on this direct final rule ends. If we receive any timely significant adverse comment, we will withdraw this final rule in part or in whole by publication of a document in the Federal Register within 30 days after the comment period ends.

ADDRESSES:

You may submit comments, identified by Docket No. 2006N-0335, by any of the following methods:

Electronic Submissions

Submit electronic comments in the following ways:

Start Printed Page 55730

Written Submissions

Submit written submissions in the following ways:

  • FAX: 301-827-6870.
  • Mail/Hand delivery/Courier [For paper, disk, or CD-ROM submissions]: Division of Dockets Management (HFA-305), Food and Drug Administration, 5630 Fishers Lane, rm. 1061, Rockville, MD 20852.

To ensure more timely processing of comments, FDA is no longer accepting comments submitted to the agency by e-mail. FDA encourages you to continue to submit electronic comments by using the Federal eRulemaking Portal or the agency Web site, as described in the Electronic Submissions portion of this paragraph.

Instructions: All submissions received must include the agency name and Docket No(s). and Regulatory Information Number (RIN) (if a RIN number has been assigned) for this rulemaking. All comments received may be posted without change to http://www.fda.gov/​ohrms/​dockets/​default.htm, including any personal information provided. For additional information on submitting comments, see the “Comments” heading of the SUPPLEMENTARY INFORMATION section of this document.

Docket: For access to the docket to read background documents or comments received, go to http://www.fda.gov/​ohrms/​dockets/​default.htm and insert the docket number(s), found in brackets in the heading of this document, into the “Search” box and follow the prompts and/or go to the Division of Dockets Management, 5630 Fishers Lane, rm. 1061, Rockville, MD 20852.

Start Further Info

FOR FURTHER INFORMATION CONTACT:

Heather S. Rosecrans, Center for Devices and Radiological Health (HFZ-404), Food and Drug Administration, 9200 Corporate Blvd., Rockville, MD 20850, 301-594-1190, ext. 143.

End Further Info End Preamble Start Supplemental Information

SUPPLEMENTARY INFORMATION:

I. What Is the Background of the Rule?

On October 26, 2002, MDUFMA (Public Law 107-250), amended the Federal Food, Drug, and Cosmetic Act (the act) by adding section 510(o) (21 U.S.C. 360(o)), which provided new regulatory requirements for reprocessed SUDs. According to this provision, in order to ensure that reprocessed SUDs are substantially equivalent to predicate devices, 510(k)s for certain reprocessed SUDs identified by FDA must include validation data. These required validation data include cleaning and sterilization data, and functional performance data demonstrating that each SUD will remain substantially equivalent to its predicate device after the maximum number of times the device is reprocessed as intended by the person submitting the premarket notification.

Before enactment of the new law, a manufacturer of a reprocessed SUD was required to obtain premarket approval or premarket clearance for the device, unless the device was exempt from premarket submission requirements. Under MDUFMA, some previously exempt reprocessed SUDs are no longer exempt from premarket notification requirements. Manufacturers of these identified devices were required to submit 510(k)s that included validation data specified by FDA. Reprocessors of certain SUDs already subject to cleared 510(k)s were also required to submit the validation data specified by the agency.

A. Definitions

Under section 201(ll)(2)(A) of the act (21 U.S.C. 321(ll)(2)(A)), a reprocessed SUD is defined as an “original device that has previously been used on a patient and has been subjected to additional processing and manufacturing for the purpose of an additional single use on a patient. The subsequent processing and manufacture of a reprocessed single-use device shall result in a device that is reprocessed within the meaning of this definition.” FDA is amending § 807.3 (21 CFR 807.3) by adding paragraph (t) to incorporate this definition into the regulations.

Reprocessed SUDs are divided into three groups: (1) Critical, (2) semicritical, and (3) noncritical. The first two categories are set forth in the act and all three reflect a categorization scheme recognized in the industry (Ref. 1). In the Federal Register of April 30, 2003 (68 FR 23139), FDA describes in more detail the development of this scheme and its use in the implementation of section 510(o) of the act. The act defines critical and semicritical reprocessed single use devices at section 201(mm) as amended by MDUFMA. FDA defined noncritical devices in the Federal Register of April 30, 2003. The definitions are as follows:

  • A critical reprocessed SUD is intended to contact normally sterile tissue or body spaces during use.
  • A semicritical reprocessed SUD is intended to contact intact mucous membranes and not penetrate normally sterile areas of the body.
  • A noncritical reprocessed SUD is intended to make topical contact and not penetrate intact skin.

B. Critical and Semicritical Reprocessed SUDs Previously Exempt from Premarket Notification

MDUFMA required FDA to review the critical and semicritical reprocessed SUDs that were previously exempt from premarket notification requirements and determine which of these devices required premarket notification to ensure their substantial equivalence to predicate devices. Under MDUFMA, FDA was required to identify in a Federal Register notice those critical reprocessed SUDs whose exemption from premarket notification would be terminated and for which FDA determined that validation data, as specified under MDUFMA, was necessary in a 510(k). FDA published a list of these devices on April 30, 2003. According to the law, manufacturers of the devices whose exemptions from premarket notification were terminated were required to submit 510(k)s that included validation data regarding cleaning, sterilization, and functional performance, in addition to all the other required elements of a 510(k) identified in § 807.87 (21 CFR 807.87), within 15 months of publication of the notice or no longer market their devices.

In accordance with section 510(o) of the act, FDA must revise the list of devices subject to this requirement as appropriate. In the Federal Register of June 26, 2003 (68 FR 38071), FDA recategorized nine device types from semicritical to critical, and added nonelectric gastroenterology-urology biopsy forceps to the list of critical reprocessed SUDs whose exemption from premarket notification requirements was being terminated. In the Federal Register of September 29, 2005 (70 FR 56911), FDA announced that it was adding devices to the list of critical reprocessed SUDs whose 510(k) exemption is terminated and for which validation data is necessary.

By April 26, 2004, FDA was required to identify in a Federal Register notice those semicritical reprocessed SUDs whose exemption from premarket notification would be terminated and for which FDA determined that validation data, as specified under MDUFMA, was necessary in a 510(k). FDA published this list in the Federal Register of April 13, 2004 (69 FR 19433). As discussed previously in this document, manufacturers of the devices whose exemptions from premarket notification were terminated were required to submit 510(k)s that included validation data regarding cleaning, sterilization, and functional performance, in addition to all the other required elements of a 510(k) identified Start Printed Page 55731in § 807.87, within 15 months of publication of the notice or no longer market their devices. In accordance with section 510(o) of the act, FDA must revise the list of devices subject to this requirement as appropriate.

C. Reprocessed SUDs Already Subject to Premarket Notification Requirements

MDUFMA also required FDA to review the types of reprocessed SUDs already subject to premarket notification requirements and to identify which of these devices required the submission of validation data to ensure their substantial equivalence to predicate devices. FDA published a list of these devices in the Federal Register of April 30, 2003. As described previously in this document, FDA must revise the list of devices subject to this requirement as appropriate. In the Federal Register of September 29, 2005, FDA announced that it was adding laparoscopic and endoscopic electrosurgical accessories to this list of reprocessed SUDs already subject to premarket notification.

For devices identified on this list that had already been cleared through the 510(k) process, manufacturers were required to submit validation data regarding cleaning, sterilization, and functional performance within 9 months of publication of the list or no longer market their devices.

For devices on this list that were not yet cleared through the 510(k) process, manufacturers were required to submit 510(k)s with validation data regarding cleaning, sterilization, and functional performance, in addition to all the other required elements identified in § 807.87.

II. What Does This Direct Final Rulemaking Do?

In this final rule, FDA is:

  • Amending § 807.3 to add definitions for “single use device,” “reprocessed single use device (SUD)” and “validation data.” The definitions of single use device and reprocessed single use device reflect the definitions of those terms in section 201(ll) of the act as amended by MDUFMA. The definition of validation data tracks the language used to describe validation data in section 510(o)(1)(A) of the act.
  • Amending § 807.87 (Information required in a premarket notification submission) to reference the requirement (of section 510(o) of the act) to submit validation data for reprocessed SUDs in appropriate situations.
  • Amending the classification regulations for devices for which FDA has revoked the exemption from premarket notification submission for reprocessed SUDs that require the submission of validation data. If the revocation applies only to a subset within a generic type, FDA has revised the language accordingly.
  • Amending the classification regulations for devices already subject to premarket notification that FDA has designated as requiring the submission of validation data. If the requirement to submit validation data applies only to a subset of the generic type, FDA has revised the regulation accordingly.
  • Making minor corrections to the classification regulations, including amending the sections for affected class II devices to update the definition of class II from requiring “performance standards” to requiring “special controls” in order to provide a reasonable assurance of the safety and effectiveness of the device. This conforms these sections to the definition of class II at section 513(a)(1)(B) of the act (21 U.S.C. 360c(a)(1)(B)), as revised by the Safe Medical Devices Act of 1990 (SMDA) (Public Law 101-629).

FDA has made available previously a guidance document on the submission of validation data entitled “Medical Device User Fee and Modernization Act of 2002, Validation Data in Premarket Notification Submissions (510(k)s) for Reprocessed Single-Use Medical Devices” (June 1, 2004, 69 FR 30943). This guidance document may be accessed on the Internet at http://www.fda.gov/​cdrh/​ode/​guidance/​1216.html.

III. What are the Procedures for Issuing a Direct Final Rule?

In the Federal Register of November 21, 1997 (62 FR 62466), FDA announced the availability of the guidance document entitled “Guidance for FDA and Industry: Direct Final Rule Procedures” that described when and how FDA will employ direct final rulemaking. We believe that this rule is appropriate for direct final rulemaking because it is intended to make noncontroversial amendments and minor corrections to existing regulations. We anticipate no significant adverse comment.

Consistent with FDA's procedures on direct final rulemaking, we are publishing elsewhere in this issue of the Federal Register a companion proposed rule that is identical to the direct final rule. The companion proposed rule provides a procedural framework within which the rule may be finalized in the event the direct final rule is withdrawn because of any significant adverse comment. The comment period for this direct final rule runs concurrently with the comment period of the companion proposed rule. Any comments received in response to the companion proposed rule will also be considered as comments regarding this direct final rule.

If we receive any significant adverse comment, we intend to withdraw this final rule before its effective date by publication of a notice in the Federal Register within 30 days after the comment period ends. A significant adverse comment is defined as a comment that explains why the rule would be inappropriate, including challenges to the rule's underlying premise or approach, or would be ineffective or unacceptable without change. In determining whether an adverse comment is significant and warrants terminating a direct final rulemaking, we will consider whether the comment raises an issue serious enough to warrant a substantive response in a notice-and-comment process in accordance with section 553 of the Administrative Procedure Act (APA) (5 U.S.C. 553). Comments that are frivolous, insubstantial, or outside the scope of the rule will not be considered significant or adverse under this procedure. For example, a comment recommending an additional change to the rule will not be considered a significant adverse comment, unless the comment states why the rule would be ineffective without the additional change. In addition, if a significant adverse comment applies to part of a rule and that part can be severed from the remainder of the rule, we may adopt as final those parts of the rule that are not the subject of a significant adverse comment.

If we withdraw the direct final rule, all comments received will be considered under the companion proposed rule in developing a final rule under the usual notice-and-comment procedures under the APA (5 U.S.C. 552a et seq.). If we receive no significant adverse comment during the specified comment period, we intend to publish a confirmation document in the Federal Register within 30 days after the comment period ends.

IV. What is the Legal Authority for This Rule?

This direct final rule is authorized by sections 201, 301, 501, 502, 510, 513, 515, 519, 520, 701, 704, 801, and 903 of the act and sections 264 and 271 of the Public Health Service Act (21 U.S.C. 321, 331, 351, 352, 360, 360c, 360e, 360i, 360j, 371, 374, 381, 393; 42 U.S.C. 264 and 271).

Start Printed Page 55732

V. What is the Environmental Impact of This Rule?

We have determined under 21 CFR 25.30(h) and 25.34(a) that this action is of a type that does not individually or cumulatively have a significant effect on the human environment. Therefore, neither an environmental assessment nor an environmental impact statement is required.

VI. What is the Economic Impact of This Rule?

We have examined the impacts of the final rule under Executive Order 12866 and the Regulatory Flexibility Act (5 U.S.C. 601-612), and the Unfunded Mandates Reform Act of 1995 (Public Law 104-4). Executive Order 12866 directs agencies to assess all costs and benefits of available regulatory alternatives and, when regulation is necessary, to select regulatory approaches that maximize net benefits (including potential economic, environmental, public health and safety, and other advantages; distributive impacts; and equity). We believe that this direct final rule is not a significant regulatory action as defined by the Executive order.

The Regulatory Flexibility Act requires agencies to analyze regulatory options that would minimize any significant impact of a rule on small entities. Because this regulation only codifies in regulations existing statutory requirements, the agency certifies that the final rule will not have a significant economic impact on a substantial number of small entities.

Section 202(a) of the Unfunded Mandates Reform Act of 1995 requires that agencies prepare a written statement, which includes an assessment of anticipated costs and benefits, before proposing “any rule that includes any Federal mandate that may result in the expenditure by State, local, and tribal governments, in the aggregate, or by the private sector, of $100,000,000 or more (adjusted annually for inflation) in any one year.” The current threshold after adjustment for inflation is $118 million, using the most current (2004) Implicit Price Deflator for the Gross Domestic Product. We do not expect this final rule to result in any 1-year expenditure that would meet or exceed this amount.

VII. How Does the Paperwork Reduction Act of 1995 Apply to This Rule?

This final rule contains information collection provisions that are subject to review by the Office of Management and Budget (OMB) under the Paperwork Reduction Act of 1995 (the PRA) (44 U.S.C. 3501-3520). The collections of information addressed in the direct final rule have been approved by OMB in accordance with the PRA under the regulations governing premarket notifications (21 CFR part 807, OMB control number 0910-0120).

VIII. What are the Federalism Impacts of This Rule?

We have analyzed this final rule in accordance with the principles set forth in Executive Order 13132. We have determined that the rule does not contain policies that have substantial direct effects on the States, on the relationship between the National Government and the States, or on the distribution of power and responsibilities among the various levels of government. Accordingly, we have concluded that the rule does not contain policies that have federalism implications as defined in the Executive order and, consequently, a federalism summary impact statement is not required.

IX. How Do You Submit Comments on This Rule?

Interested persons may submit to the Division of Dockets Management (see ADDRESSES) written or electronic comments regarding this direct final rule. Submit a single copy of electronic comments or two paper copies of any mailed comments, except that individuals may submit one paper copy. Comments are to be identified with the docket number found in brackets in the heading of this document. Received comments may be seen in the Division of Dockets Management between 9 a.m. and 4 p.m., Monday through Friday.

X. What Is the Reference for This Rule?

The following reference has been placed on display in the Division of Dockets Management (see ADDRESSES) and may be seen by interested persons between 9 am. and 4 pm., Monday through Friday.

1. Spaulding, E. H., “The Role of Chemical Disinfection in the Prevention of Nonsocomial Infections.” Edited by P. S. Brachman and T. C. Eickof, Proceedings of International Conference on Nonsocomial Infections, 1970, American Hospital Association, Chicago, 254-274, 1971.

Start List of Subjects

List of Subjects

, 870, 872, 874, 876, 878, 880, 882, and 884

End List of Subjects Start Amendment Part

Therefore, under the Federal Food, Drug, and Cosmetic Act and under authority delegated to the Commissioner of Food and Drugs,

End Amendment Part Start Part

PART 807—ESTABLISHMENT REGISTRATION AND DEVICE LISTING FOR MANUFACTURERS AND INITIAL IMPORTERS OF DEVICES

End Part Start Amendment Part

1. The authority citation for

End Amendment Part Start Authority

Authority: 21 U.S.C. 321, 331, 351, 352, 360, 360c, 360e, 360i, 360j, 371, 374, 381, 393; 42 U.S.C. 264, 271.

End Authority Start Amendment Part

2. Section 807.3 is amended by adding new paragraphs (t), (u), and (v) to read as follows:

End Amendment Part
Definitions.
* * * * *

(t) A single use device (SUD) means a device that is intended for one use or on a single patient during a single procedure.

(u) A reprocessed SUD is an original device that has previously been used on a patient and has been subjected to additional processing and manufacturing for the purpose of an additional single use on a patient. The subsequent processing and manufacture of a reprocessed SUD shall result in a device that is reprocessed within the meaning of this definition.

(v) Validation data for the purposes of this part means cleaning and sterilization data, and functional performance data demonstrating that an SUD will remain substantially equivalent to its predicate device after the maximum number of times the device is reprocessed as intended by the person submitting the premarket notification.

Start Amendment Part

3. Section 807.87 is amended by redesignating paragraphs (h), (i), (j), (k), and (l) as paragraphs (i), (j), (k), (l), and (m), respectively, and by adding new paragraph (h) to read as follows:

End Amendment Part
Information required in a premarket notification submission.
* * * * *

(h) If the device is a reprocessed SUD that FDA has identified as requiring validation data, the premarket Start Printed Page 55733notification submission must include validation data as defined in § 807.3(v).

* * * * *
Start Part

PART 868—ANESTHESIOLOGY DEVICES

End Part Start Amendment Part

4. The authority citation for

End Amendment Part Start Authority

Authority: 21 U.S.C. 351, 360, 360c, 360e, 360j, 371.

End Authority Start Amendment Part

5. Section 868.5150 is amended by revising paragraph (b) to read as follows:

End Amendment Part
Anesthesia conduction needle.
* * * * *

(b) Classification. Class II (special controls). If the device is an anesthetic conduction needle (with/without introducer) or a short term spinal needle and it is a reprocessed single use device (SUD) as defined in § 807.3(u) of this chapter, a premarket notification submission for the device must include validation data as described in § 807.3(v).

Start Amendment Part

6. Section 868.5730 is amended by revising paragraph (b) to read as follows:

End Amendment Part
Tracheal tube.
* * * * *

(b) Classification. Class II (special controls). If the device is a reprocessed single use device (SUD) as defined in § 807.3(u) of this chapter, a premarket notification submission for the device must include validation data as described in § 807.3(v).

Start Amendment Part

7. Section 868.5905 is amended by revising paragraph (b) to read as follows:

End Amendment Part
Noncontinuous ventilator (IPPB).
* * * * *

(b) Classification. Class II (special controls). If the device is a noncontinuous ventilator (respirator) mask that is a reprocessed single use device (SUD) as defined in § 807.3(u) of this chapter, a premarket notification submission for the device must include validation data as described in § 807.3(v).

Start Amendment Part

8. Section 868.6810 is amended by revising paragraph (b) to read as follows:

End Amendment Part
Tracheobronchial suction catheter.
* * * * *

(b) Classification. Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 868.9. If the device is a reprocessed single use device (SUD) as defined in § 807.3(u) of this chapter, the exemption from premarket notification does not apply and the manufacturer must submit a premarket notification that includes validation data as described in § 807.3(v).

Start Part

PART 870—CARDIOVASCULAR DEVICES

End Part Start Amendment Part

9. The authority citation for

End Amendment Part Start Authority

Authority: 21 U.S.C. 351, 360, 360c, 360e, 360j, 371.

End Authority Start Amendment Part

10. Section 870.1200 is amended by revising paragraph (b) to read as follows:

End Amendment Part
Diagnostic intravascular catheter.
* * * * *

(b) Classification. Class II (special controls). If the device is an angiography catheter that is a reprocessed single use device (SUD) as defined in § 807.3(u) of this chapter, a premarket notification submission for the device must include validation data as described in § 807.3(v).

Start Amendment Part

11. Section 870.1220 is amended by revising paragraph (b) to read as follows:

End Amendment Part
Electrode recording catheter or electrode recording probe.
* * * * *

(b) Classification. Class II (special controls). If the device is an electrode recording catheter or intracardiac mapping catheter that is a reprocessed single use device (SUD) as defined in § 807.3(u) of this chapter, a premarket notification submission for the device must include validation data as described in § 807.3(v).

Start Amendment Part

12. Section 870.1230 is amended by revising paragraph (b) to read as follows:

End Amendment Part
Fiberoptic oximeter catheter.
* * * * *

(b) Classification. Class II (special controls). If the device is a reprocessed single use device (SUD) as defined in § 807.3(u) of this chapter, a premarket notification submission for the device must include validation data as described in § 807.3(v).

Start Amendment Part

13. Section 870.1280 is amended by revising paragraph (b) to read as follows:

End Amendment Part
Steerable catheter.
* * * * *

(b) Classification. Class II (special controls). If the device is a reprocessed single use device (SUD) as defined in § 807.3(u) of this chapter, a premarket notification submission for the device must include validation data as described in § 807.3(v).

Start Amendment Part

14. Section 870.1290 is amended by revising paragraph (b) to read as follows:

End Amendment Part
Steerable catheter control system.
* * * * *

(b) Classification. Class II (special controls). If the device is a reprocessed single use device (SUD) as defined in § 807.3(u) of this chapter, a premarket notification submission for the device must include validation data as described in § 807.3(v).

Start Amendment Part

15. Section 870.1330 is amended by revising paragraph (b) to read as follows:

End Amendment Part
Catheter guide wire.
* * * * *

(b) Classification. Class II (special controls). If the device is a reprocessed single use device (SUD) as defined in § 807.3(u) of this chapter, a premarket notification submission for the device must include validation data as described in § 807.3(v).

Start Amendment Part

16. Section 870.1390 is amended by revising paragraph (b) to read as follows:

End Amendment Part
Trocar.
* * * * *

(b) Classification. Class II (special controls). If the device is a cardiovascular trocar that is a reprocessed single use device (SUD) as defined in § 807.3(u) of this chapter, a premarket notification submission for the device must include validation data as described in § 807.3(v).

17. Section 870.1650 is amended by revising paragraph (b) to read as follows:

Angiographic injector and syringe.
* * * * *

(b) Classification. Class II (special controls). If the device is a reprocessed single use device (SUD) as defined in § 807.3(u) of this chapter, a premarket notification submission for the device must include validation data as described in § 807.3(v).

Start Amendment Part

18. Section 870.1670 is amended by revising paragraph (b) to read as follows:

End Amendment Part
Syringe actuator for an injector.
* * * * *

(b) Classification. Class II (special controls). If the device is a reprocessed single use device (SUD) as defined in § 807.3(u) of this chapter, a premarket notification submission for the device must include validation data as described in § 807.3(v).

Start Amendment Part

19. Section 870.2700 is amended by revising paragraph (b) to read as follows:

End Amendment Part
Oximeter.
* * * * *

(b) Classification. Class II (special controls). If the device is a tissue saturation oximeter or an oximeter and it is a reprocessed single use device (SUD) as defined in § 807.3(u) of this chapter, a premarket notification submission for the device must include Start Printed Page 55734validation data as described in § 807.3(v).

Start Amendment Part

20. Section 870.3535 is amended by revising paragraph (b) to read as follows:

End Amendment Part
Intra-aortic balloon and control system.
* * * * *

(b) Classification. Class III (premarket approval). If the device is a reprocessed single use device (SUD) as defined in § 807.3(u) of this chapter, a premarket notification submission for the device must include validation data as described in § 807.3(v).

* * * * *
Start Amendment Part

21. Section 870.4450 is amended by revising paragraph (b) to read as follows:

End Amendment Part
Vascular clamp.
* * * * *

(b) Classification. Class II (special controls). If the device is a reprocessed single use device (SUD) as defined in § 807.3(u) of this chapter, a premarket notification submission for the device must include validation data as described in § 807.3(v).

Start Amendment Part

22. Section 870.4500 is amended by revising paragraph (b) to read as follows:

End Amendment Part
Cardiovascular surgical instruments.
* * * * *

(b) Classification. Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9. If the device is a noncompression heart stabilizer that is a reprocessed single use device (SUD) as defined in § 807.3(u) of this chapter, the exemption from premarket notification does not apply and the manufacturer must submit a premarket notification that includes validation data as described in § 807.3(v).

Start Amendment Part

23. Section 870.4885 is amended by revising paragraph (b) to read as follows:

End Amendment Part
External vein stripper.
* * * * *

(b) Classification. Class II (special controls). If the device is a reprocessed single use device (SUD) as defined in § 807.3(u) of this chapter, a premarket notification submission for the device must include validation data as described in § 807.3(v).

Start Part

PART 872—DENTAL DEVICES

End Part Start Amendment Part

24. The authority citation for

End Amendment Part Start Authority

Authority: 21 U.S.C. 351, 360, 360c, 360e, 360j, 371.

End Authority Start Amendment Part

25. Section 872.3240 is amended by revising paragraph (b) to read as follows:

End Amendment Part
Dental bur.
* * * * *

(b) Classification. Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 872.9. If the device is a dental diamond coated bur that is a reprocessed single use device (SUD) as defined in § 807.3(u) of this chapter, the exemption from premarket notification does not apply and the manufacturer must submit a premarket notification submission that includes validation data as described in § 807.3(v).

Start Amendment Part

26. Section 872.4535 is amended by revising paragraph (b) to read as follows:

End Amendment Part
Dental diamond instrument.
* * * * *

(b) Classification. Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 872.9. If the device is a reprocessed single use device (SUD) as defined in § 807.3(u) of this chapter, the exemption from premarket notification does not apply and the manufacturer must submit a premarket notification that includes validation data as described in § 807.3(v).

Start Amendment Part

27. Section 872.4730 is amended by revising paragraph (b) to read as follows:

End Amendment Part
Dental injecting needle.
* * * * *

(b) Classification. Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 872.9. If the device is a reprocessed single use device (SUD) as defined in § 807.3(u) of this chapter, the exemption from premarket notification does not apply and the manufacturer must submit a premarket notification that includes validation data as described in § 807.3(v).

Start Amendment Part

28. Section 872.5410 is amended by revising paragraph (b) to read as follows:

End Amendment Part
Orthodontic appliance and accessories.
* * * * *

(b) Classification. Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 872.9. If the device is a orthodontic metal bracket that is a reprocessed single use device (SUD) as defined in § 807.3(u) of this chapter, the exemption from premarket notification does not apply and the manufacturer must submit a premarket notification that includes validation data as described in § 807.3(v).

Start Amendment Part

29. Section 872.5470 is amended by revising paragraph (b) to read as follows:

End Amendment Part
Orthodontic plastic bracket.
* * * * *

(b) Classification. Class II (special controls). If the device is a reprocessed single use device (SUD) as defined in § 807.3(u) of this chapter, the exemption from premarket notification does not apply and the manufacturer must submit a premarket notification that includes validation data as described in § 807.3(v).

Start Part

PART 874—EAR, NOSE, AND THROAT DEVICES

End Part Start Amendment Part

30. The authority citation for

End Amendment Part Start Authority

Authority: 21 U.S.C. 351, 360, 360c, 360e, 360j, 371.

End Authority Start Amendment Part

31. Section 874.4140 is amended by revising paragraph (b) to read as follows:

End Amendment Part
Ear, nose, and throat bur.
* * * * *

(b) Classification. Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 874.9. If the device is an ear, nose, and throat (ENT) high speed microdebrider or an ENT diamond coated bur and it is a reprocessed single use device (SUD) as defined in § 807.3(u) of this chapter, the exemption from premarket notification does not apply and the manufacturer must submit a premarket notification that includes validation data as described in § 807.3(v).

Start Amendment Part

32. Section 874.4420 is amended by revising paragraph (b) to read as follows:

End Amendment Part
Ear, nose, and throat manual surgical instrument.
* * * * *

(b) Classification. Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 874.9. If the device is a laryngeal, sinus, or tracheal trocar that is a reprocessed single use device (SUD) as defined in § 807.3(u) of this chapter, the exemption from premarket notification does not apply and the manufacturer must submit a premarket notification that includes validation data as described in § 807.3(v).

Start Amendment Part

33. Section 874.4680 is amended by revising paragraph (b) to read as follows:

End Amendment Part
Start Printed Page 55735
Bronchoscope (flexible or rigid) and accessories.
* * * * *

(b) Classification. Class II (special controls). If the device is a bronchoscope (nonrigid) biopsy forceps that is a reprocessed single use device (SUD) as defined in § 807.3(u) of this chapter, a premarket notification submission for the device must include validation data as described in § 807.3(v).

Start Part

PART 876—GASTROENTEROLOGY-UROLOGY DEVICES

End Part Start Amendment Part

34. The authority citation for

End Amendment Part Start Authority

Authority: 21 U.S.C. 351, 360, 360c, 360e, 360j, 360l, 371.

End Authority Start Amendment Part

35. Section 876.1075 is amended by revising paragraph (b) to read as follows:

End Amendment Part
Gastroenterology-urology biopsy instrument.
* * * * *

(b) Classification. (1) Class II (special controls). If the device is a gastroenterology-urology (G-U) biopsy needle and needle set or a biopsy instrument and it is a reprocessed single use device (SUD) as defined in § 807.3(u) of this chapter, a premarket notification submission for the device must include validation data as described in § 807.3(v).

(2) Class I (general controls) for the biopsy forceps cover and the non-electric biopsy forceps. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9. If the device is a non-electric biopsy forceps that is a reprocessed SUD as defined in § 807.3(u) of this chapter, the exemption from premarket notification does not apply and the manufacturer must submit a premarket notification that includes validation data as described in § 807.3(v).

Start Amendment Part

36. Section 876.1500 is amended by revising paragraph (b) (1) to read as follows:

End Amendment Part
Endoscope and accessories.
* * * * *

(b) Classification. (1) Class II (special controls). If the device is an endoscopic needle, an endoilluminator, a general and plastic surgery laparoscope, or a spring-loaded pneumoperitoneum needle and it is a reprocessed single use device (SUD) as defined in § 807.3(u) of this chapter, a premarket notification submission for the device must include validation data as described in § 807.3(v).

* * * * *
Start Amendment Part

37. Section 876.4300 is amended by revising paragraph (b) to read as follows:

End Amendment Part
Endoscopic electrosurgical unit and accessories.
* * * * *

(b) Classification. Class II (special controls). If the device is an active urological electrosurgical electrode, a flexible suction coagulator electrode, an electric biopsy forceps, a flexible snare, or an endoscopic (with or without accessories) electrosurgical unit that is a reprocessed single use device (SUD) as defined in § 807.3(u) of this chapter, a premarket notification submission for the device must include validation data as described in § 807.3(v).

Start Amendment Part

38. Section 876.4680 is amended by revising paragraph (b) to read as follows:

End Amendment Part
Ureteral stone dislodger.
* * * * *

(b) Classification. Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9. If the device is a flexible and basket stone dislodger that is a reprocessed single use device (SUD) as defined in § 807.3(u) of this chapter, a premarket notification submission for the device must include validation data as described in § 807.3(v).

Start Amendment Part

39. Section 876.5010 is amended by revising paragraph (b) to read as follows:

End Amendment Part
Biliary catheter and accessories.
* * * * *

(b) Classification. Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9. If the device is a biliary catheter that is a reprocessed single use device (SUD) as defined in § 807.3(u) of this chapter, a premarket notification submission for the device must include validation data as described in § 807.3(v).

Start Amendment Part

40. Section 876.5540 is amended in paragraph (b) (3) to read as follows:

End Amendment Part
Blood access device and accessories.
* * * * *

(b) * * *

(3) Class II (special controls) for accessories for both the implanted and the nonimplanted blood access devices not listed in paragraph (b)(4) of this section. If the device is a single needle dialysis set (coaxial flow) or fistula needle and it is a reprocessed single use device (SUD) as defined in § 807.3(u) of this chapter, a premarket notification submission for the device must include validation data as described in § 807.3(v).

* * * * *
Start Amendment Part

41. Section 876.5820 is amended by revising paragraph (b) (1) to read as follows:

End Amendment Part
Hemodialysis system and accessories.
* * * * *

(b) Classification. (1) Class II (special controls) (for hemodialysis systems and all accessories directly associated with the extracorporeal blood system and the dialysate delivery system). If the device is a single needle dialysis set with uni-directional pump that is a reprocessed single use device (SUD) as defined in § 807.3(u) of this chapter, a premarket notification submission for the device must include validation data as described in § 807.3(v).

* * * * *
Start Part

PART 878—GENERAL AND PLASTIC SURGERY DEVICES

End Part Start Amendment Part

42. The authority citation for

End Amendment Part Start Authority

Authority: 21 U.S.C. 351, 360, 360c, 360e, 360j, 360l, 371.

End Authority Start Amendment Part

43. Section 878.4200 is amended by revising paragraph (b) to read as follows:

End Amendment Part
Introduction/drainage catheter and accessories.
* * * * *

(b) Classification. Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 878.9. If the device is a catheter needle that is a reprocessed single use device (SUD) as defined in § 807.3(u) of this chapter, the exemption from premarket notification does not apply and the manufacturer must submit a premarket notification that includes validation data as described in § 807.3(v).

Start Amendment Part

44. Section 878.4300 is amended by revising paragraph (b) to read as follows:

End Amendment Part
Implantable clip.
* * * * *

(b) Classification. Class II (special controls). If the device is a reprocessed single use device (SUD) as defined in § 807.3(u) of this chapter, a premarket notification submission for the device must include validation data as described in § 807.3(v).

Start Amendment Part

45. Section 878.4400 is amended by revising paragraph (b) to read as follows:

End Amendment Part
Electrosurgical cutting and coagulation device and accessories.
* * * * *
Start Printed Page 55736

(b) Classification. Class II (special controls). If the device is an endoscopic or laparoscopic electrosurgical accessory that is a reprocessed single use device (SUD) as defined in § 807.3(u) of this chapter, a premarket notification submission for the device must include validation data as described in § 807.3(v).

Start Amendment Part

46. Section 878.4750 is amended by revising paragraph (b) to read as follows:

End Amendment Part
Implantable staple.
* * * * *

(b) Classification. Class II (special controls). If the device is a reprocessed single use device (SUD) as defined in § 807.3(u) of this chapter, a premarket notification submission for the device must include validation data as described in § 807.3(v).

Start Amendment Part

47. Section 878.4800 is amended by revising paragraph (b) to read as follows:

End Amendment Part
Manual surgical instrument for general use.
* * * * *

(b) Classification. Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 878.9. If the device is a percutaneous biopsy device, a gastroenterology-urology needle, a cardiovascular biopsy needle, or an aspiration and injection needle and it is a reprocessed single use device (SUD) as defined in § 807.3(u) of this chapter, the exemption from premarket notification does not apply and the manufacturer must submit a premarket notification that includes validation data as described in § 807.3(v).

Start Part

PART 880—GENERAL HOSPITAL AND PERSONAL USE DEVICES

End Part Start Amendment Part

48. The authority citation for

End Amendment Part Start Authority

Authority: 21 U.S.C. 351, 360, 360c, 360e, 360j, 371.

End Authority Start Amendment Part

49. Section 880.5570 is amended by revising paragraph (b) to read as follows:

End Amendment Part
Hypodermic single lumen needle.
* * * * *

(b) Classification. Class II (special controls). If the device is a reprocessed single use device (SUD) as defined in § 807.3(u) of this chapter, a premarket notification submission for the device must include validation data as described in § 807.3(v).

Start Amendment Part

50. Section 880.5860 is amended by revising paragraph (b) to read as follows:

End Amendment Part
Piston syringe.
* * * * *

(b) Classification. Class II (special controls). If the device is a reprocessed single use device (SUD) as defined in § 807.3(u) of this chapter, a premarket notification submission for the device must include validation data as described in § 807.3(v).

Start Part

PART 882—NEUROLOGICAL DEVICES

End Part Start Amendment Part

51. The authority citation for

End Amendment Part Start Authority

Authority: 21 U.S.C. 351, 360, 360c, 360e, 360j, 371.

End Authority Start Amendment Part

52. Section 882.4190 is amended by revising paragraph (b) to read as follows:

End Amendment Part
Clip forming/cutting instrument.
* * * * *

(b) Classification. Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 882.9. If the device is a reprocessed single use device (SUD) as defined in § 807.3(u) of this chapter, the exemption from premarket notification does not apply and the manufacturer must submit a premarket notification that includes validation data as described in § 807.3(v).

Start Amendment Part

53. Section 882.4300 is amended by revising paragraph (b) to read as follows:

End Amendment Part
Manual cranial drills, burrs, trephines, and accessories.
* * * * *

(b) Classification. Class II (special controls). If the device is a reprocessed single use device (SUD) as defined in § 807.3(u) of this chapter, a premarket notification submission for the device must include validation data as described in § 807.3(v).

Start Amendment Part

54. Section 882.4305 is amended by revising paragraph (b) to read as follows:

End Amendment Part
Powered compound cranial drills, burrs, trephines, and their accessories.
* * * * *

(b) Classification. Class II (special controls). If the device is a reprocessed single use device (SUD) as defined in § 807.3(u) of this chapter, a premarket notification submission for the device must include validation data as described in § 807.3(v).

Start Amendment Part

55. Section 882.4310 is amended by revising paragraph (b) to read as follows:

End Amendment Part
Powered simple cranial drills, burrs, trephines, and their accessories.
* * * * *

(b) Classification. Class II (special controls). If the device is a reprocessed single use device (SUD) as defined in § 807.3(u) of this chapter, a premarket notification submission for the device must include validation data as described in § 807.3(v).

Start Part

PART 884—OBSTETRICAL AND GYNECOLOGICAL DEVICES

End Part Start Amendment Part

56. The authority citation for

End Amendment Part Start Authority

Authority: 21 U.S.C. 351, 360, 360c, 360e, 360j, 371.

End Authority Start Amendment Part

57. Section 884.1720 is amended by revising paragraph (b)(1) to read as follows:

End Amendment Part
Gynecologic laparoscope and accessories.
* * * * *

(b) (1) Classification. Class II (special controls). If the device is a reprocessed single use device (SUD) as defined in § 807.3(u) of this chapter, a premarket notification submission for the device must include validation data as described in § 807.3(v).

* * * * *
Start Amendment Part

58. Section 884.1730 is amended by revising paragraph (b)(2) to read as follows:

End Amendment Part
Laparoscopic insufflator.
* * * * *

(b) * * *

(2) Class I for tubing and tubing/filter kits which include accessory instruments that are not used to effect intra-abdominal insufflation (pneumoperitoneum). The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 884.9. If the device is a reprocessed single use device (SUD) as defined in § 807.3(u) of this chapter, the exemption from premarket notification does not apply and the manufacturer must submit a premarket notification that includes validation data as described in § 807.3(v).

Start Amendment Part

59. Section 884.4530 is amended by revising paragraph (b)(2) to read as follows:

End Amendment Part
Obstetric-gynecologic specialized manual instrument.
* * * * *

(b) * * *

(2) Class I for the amniotome, uterine curette, cervical dilator (fixed-size bougies), cerclage needle, intrauterine device (IUD) remover, uterine sound, and gynecological biopsy forceps. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 884.9. If the device is a Start Printed Page 55737gynecological biopsy forceps that is a reprocessed single use device (SUD) as defined in § 807.3(u) of this chapter, the exemption from premarket notification does not apply and the manufacturer must submit a premarket notification that includes validation data as described in § 807.3(v).

Start Amendment Part

60. Section 884.6100 is amended by revising paragraph (b) to read as follows:

End Amendment Part
Assisted reproduction needles.
* * * * *

(b) Classification. Class II (special controls) (mouse embryo assay information, endotoxin testing, sterilization validation, design specifications, labeling requirements, biocompatibility testing, and clinical testing). If the device is a reprocessed single use device (SUD) as defined in § 807.3(u) of this chapter, the exemption from premarket notification does not apply and the manufacturer must submit a premarket notification that includes validation data as described in § 807.3(v).

Start Part

PART 886—OPTHALMIC DEVICES

End Part Start Amendment Part

61. The authority citation for

End Amendment Part Start Authority

Authority: 21 U.S.C. 351, 360, 360c, 360e, 360j, 371.

End Authority Start Amendment Part

62. Section 886.4350 is amended by revising paragraph (b) to read as follows:

End Amendment Part
Manual ophthalmic surgical instrument .
* * * * *

(b) Classification. Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 886.9. If the device is an ophthalmic knife that is a reprocessed single use device (SUD) as defined in § 807.3(u) of this chapter, the exemption from premarket notification does not apply and the manufacturer must submit a premarket notification that includes validation data as described in § 807.3(v).

Start Amendment Part

63. Section 886.4370 is amended by revising paragraph (b) to read as follows:

End Amendment Part
Keratome.
* * * * *

(b) Classification. Class I (general controls). If the device is a reprocessed single use device (SUD) as defined in § 807.3(u) of this chapter, a premarket notification submission for the device must include validation data as described in § 807.3(v).

Start Amendment Part

64. Section 886.4670 is amended by revising paragraph (b) to read as follows:

End Amendment Part
Phacofragmentation system.
* * * * *

(b) Classification. Class II (special controls). If the device is a phacoemulsification needle that is a reprocessed single use device (SUD) as defined in § 807.3(u) of this chapter, a premarket notification submission for the device must include validation data as described in § 807.3(v).

Start Part

PART 892—RADIOLOGY DEVICES

End Part Start Amendment Part

65. The authority citation for

End Amendment Part Start Authority

Authority: 21 U.S.C. 351, 360, 360c, 360e, 360j, 371.

End Authority Start Amendment Part

66. Section 892.5730 is amended by revising paragraph (b) to read as follows:

End Amendment Part
Radionuclide brachytherapy source.
* * * * *

(b) Classification. Class II (special controls). If the device is an isotope needle that is a reprocessed single use device (SUD) as defined in § 807.3(u) of this chapter, a premarket notification submission for the device must include validation data as described in § 807.3(v).

Start Signature

Dated: September 6, 2006.

Jeffrey Shuren,

Assistant Commissioner for Policy.

End Signature End Supplemental Information