This site displays a prototype of a “Web 2.0” version of the daily
Federal Register. It is not an official legal edition of the Federal
Register, and does not replace the official print version or the official
electronic version on GPO’s govinfo.gov.
The documents posted on this site are XML renditions of published Federal
Register documents. Each document posted on the site includes a link to the
corresponding official PDF file on govinfo.gov. This prototype edition of the
daily Federal Register on FederalRegister.gov will remain an unofficial
informational resource until the Administrative Committee of the Federal
Register (ACFR) issues a regulation granting it official legal status.
For complete information about, and access to, our official publications
and services, go to
About the Federal Register
on NARA's archives.gov.
The OFR/GPO partnership is committed to presenting accurate and reliable
regulatory information on FederalRegister.gov with the objective of
establishing the XML-based Federal Register as an ACFR-sanctioned
publication in the future. While every effort has been made to ensure that
the material on FederalRegister.gov is accurately displayed, consistent with
the official SGML-based PDF version on govinfo.gov, those relying on it for
legal research should verify their results against an official edition of
the Federal Register. Until the ACFR grants it official status, the XML
rendition of the daily Federal Register on FederalRegister.gov does not
provide legal notice to the public or judicial notice to the courts.
Notice
Enter a search term or FR citation e.g.
88 FR 38230 FR 78782024-13208USDA09/05/24RULE0503-AA39SORN
Choosing an item from
full text search results
will bring you to those results. Pressing enter in the search box
will also bring you to search results.
Choosing an item from
suggestions
will bring you directly to the content.
Blood Glucose Monitoring Test Systems for Prescription Point-of-Care Use; Draft Guidance for Industry and Food and Drug Administration Staff; Availability
Although you can comment on any guidance at any time (see 21 CFR 10.115(g)(5)), to ensure that the Agency considers your comment on this draft guidance before it begins work on the final version of the guidance, submit either electronic or written comments on the draft guidance by April 7, 2014.
Table of Contents
Enhanced Content - Table of Contents
This table of contents is a navigational tool, processed from the
headings within the legal text of Federal Register documents.
This repetition of headings to form internal navigation links
has no substantive legal effect.
Blood Glucose Monitoring Test Systems for Prescription Point-of-Care Use; Draft Guidance for Industry and Food and Drug Administration Staff; Availability
Public Comments
Enhanced Content - Public Comments
This feature is not available for this document.
Regulations.gov Data
Enhanced Content - Regulations.gov Data
Additional information is not currently available for this document.
Document page views are updated periodically throughout the day and are
cumulative counts for this document. Counts are subject to sampling,
reprocessing and revision (up or down) throughout the day.
Page views
2,874
as of
10/15/2026 at 2:15 pm EDT
Other Formats
Enhanced Content - Other Formats
This document is also available in the following formats:
This PDF is FR Doc. 2014-00023 as it appeared on Public Inspection on
01/06/2014 at 8:45 am.
If you are using public inspection listings for legal research, you
should verify the contents of the documents against a final, official
edition of the Federal Register. Only official editions of the
Federal Register provide legal notice of publication to the public and judicial notice
to the courts under 44 U.S.C. 1503 & 1507.
Learn more here.
Published Document: 2014-00023 (79 FR 830)
This document has been published in the Federal Register. Use the PDF linked in the document sidebar for the official electronic format.
AGENCY:
Food and Drug Administration, HHS.
ACTION:
Notice.
SUMMARY:
The Food and Drug Administration (FDA) is announcing the availability of the draft guidance entitled “Blood Glucose Monitoring Test Systems for Prescription Point-of-Care Use.” This draft guidance document describes studies and criteria FDA recommends for blood glucose monitoring test systems (BGMSs) which are for prescription point-of-care use. When finalized, FDA intends for this document to guide manufacturers in conducting appropriate performance studies and preparing premarket notifications for these device types. This draft guidance is not final nor is it in effect at this time.
DATES:
Although you can comment on any guidance at any time (see 21 CFR 10.115(g)(5)), to ensure that the Agency considers your comment on this draft guidance before it begins work on the final version of the guidance, submit either electronic or written comments on the draft guidance by April 7, 2014.
ADDRESSES:
Submit written requests for single copies of the draft guidance document entitled “Blood Glucose Monitoring Test Systems for Prescription Point-of-Care Use ” to the Division of Small Manufacturers, International, and Consumer Assistance, Center for Devices and Radiological Health, Food and Drug Administration, 10903 New Hampshire Ave., Bldg. 66, rm. 4613, Silver Spring, MD 20993-0002. Send one self-addressed adhesive label to assist that office in processing your request, or fax your request to 301-847-8149. See the
SUPPLEMENTARY INFORMATION
section for information on electronic access to the guidance.
Submit electronic comments on the draft guidance to
http://www.regulations.gov.
Submit written comments to the Division of Dockets Management (HFA-305), Food and Drug Administration, 5630 Fishers Lane, rm. 1061, Rockville, MD 20852. Identify comments with the docket number found in brackets in the heading of this document.
FOR FURTHER INFORMATION CONTACT:
Patricia Bernhardt, Center for Devices and Radiological Health, Food and Drug Administration, 10903 New Hampshire Ave., Bldg. 66, rm. 5654, Silver Spring, MD 20993-0002, 301-796-6136.
SUPPLEMENTARY INFORMATION:
I. Background
This draft guidance document describes studies and criteria FDA recommends for blood glucose monitoring test systems (BGMSs) which are for prescription point-of-care use. When finalized, FDA intends for this document to guide manufacturers in conducting appropriate performance studies and preparing premarket notifications for these device types. Portable blood glucose monitoring test systems (glucose meters) that measure blood glucose concentrations are widely used in hospitals as well as in a variety of other clinical settings including both acute and chronic care facilities, general hospital wards and intensive care units, physicians' offices, assisted living facilities and nursing homes.
Historically, FDA has not recommended different types of information in premarket submissions (510(k)s) for blood glucose meters used by medical professionals as compared to over-the-counter self-monitoring devices intended for use by lay users. In recent years, however, concerns have been raised including infection control issues related to point-of-care glucose meters. According to the Centers for Medicare and Medicaid Services (CMS) and the Centers for Disease Control and Prevention (CDC), blood glucose monitoring devices can transmit bloodborne pathogens if these devices are contaminated with blood specimens and are shared between users without effective cleaning, disinfecting and appropriate infection control measures. Because BGMS devices, which are used in professional healthcare settings, are more likely to be used on multiple patients, this type of use requires certain design features and cleaning capability to prevent the spread of blood-borne pathogens.
In addition, concerns have been raised citing the inability of currently cleared BGMS devices to perform effectively in professional healthcare settings because the device's safety and effectiveness have not been evaluated for some of the intended use populations. Patients in these settings are often fundamentally different than lay users using these devices at home. Patients in professional healthcare settings can be acutely ill and medically fragile and are more likely to present physiological and pathological factors
( printed page 831)
that could interfere with glucose measurements as compared to the lay population. Errors in BGMS device accuracy can lead to incorrect insulin dosing, which, when combined with other factors, can lead to increased episodes of hypoglycemia. For hospitalized patients who may be seriously ill, any inaccuracies in the meters would further increase the risk to these patients. Previously, most blood glucose monitoring devices, even those intended to be used by healthcare professionals, were submitted to FDA with claims for OTC use. Thus, they were evaluated for use in the lay population, and the specific issues that occur in the professional healthcare setting were never addressed, the performance of the devices was not evaluated in the intended use population, and the scientific and clinical issues may not have been adequately addressed for these uses. Therefore, where devices are intended for use in professional healthcare settings, distinct performance parameters are proposed as recommendations in the draft guidance.
II. Significance of Guidance
This draft guidance is being issued consistent with FDA's good guidance practices regulation (21 CFR 10.115). The draft guidance, when finalized, will represent the Agency's current thinking on Blood Glucose Monitoring Test Systems for Prescription Point-of-Care Use . It does not create or confer any rights for or on any person and does not operate to bind FDA or the public. An alternative approach may be used if such approach satisfies the requirements of the applicable statute and regulations.
III. Electronic Access
Persons interested in obtaining a copy of the draft guidance may do so by using the Internet. A search capability for all CDRH guidance documents is available at
http://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/GuidanceDocuments/default.htm.
Guidance documents are also available at
http://www.regulations.gov.
To receive “Blood Glucose Monitoring Test Systems for Prescription Point-of-Care Use”, you may either send an email request to
dsmica@fda.hhs.gov
to receive an electronic copy of the document or send a fax request to 301-847-8149 to receive a hard copy. Please use the document number 1755 to identify the guidance you are requesting.
IV. Paperwork Reduction Act of 1995
This draft guidance refers to previously approved collections of information found in FDA regulations. These collections of information are subject to review by the Office of Management and Budget (OMB) under the Paperwork Reduction Act of 1995 (44 U.S.C. 3501-3520).The collections of information in 21 CFR part 807 subpart E have been approved under OMB control number 0910-0120; the collections of information in 21 CFR 801 and 21 CFR 809.10 have been approved under OMB control number 0910-0485; the collections of information in 21 CFR part 820 have been approved under OMB control number 0910-0073.
V. Comments
Interested persons may submit either electronic comments regarding this document to
http://www.regulations.gov
or written comments to the Division of Dockets Management (see
ADDRESSES
). It is only necessary to send one set of comments. Identify comments with the docket number found in brackets in the heading of this document. Received comments may be seen in the Division of Dockets Management between 9 a.m. and 4 p.m., Monday through Friday, and will be posted to the docket at
http://www.regulations.gov.